Somatic Vision Alive iFeel User Manual Supplement Download Page 9

THE IFEEL LABS DEVICE IS CONNECTED BUT IS SENDING BAD DATA 

If you are able to connect to the iFeel device, but you get bad data please check the Port Settings for your 
COM port. Open Device Manager (click Start, then type "device manager", then click Device Manager) then 
double click the "Standard Serial over Bluetooth" port(s) that the iFeel device is using. Click on the Port 
Settings tab and ensure the settings are Bits per second: 9600, Data bits: 8, Parity: None, Stop bits: 1, Flow 
control: None. 

If you change Port Settings please write down the prior values in case you accidentally modify port 

settings for a different device and need to revert the changes. 

 

I OWN ALIVE PIONEER WITH A GP8 AMP – HOW CAN I USE THE IFEEL SENSOR WITH THE GP8 AMP

 

From the Alive Main Menu open the Heart Rate Source dropdown and select BVP – iFeel Bluetooth. 

THE HEART RATE LINE STAYS RED (OR MOSTLY RED) IN ALIVE 

– OR –   FEW OR NO BEATS ARE DETECTED 

Please first check your sensors, then check the Bluetooth connection and adapter. 

CHECK SENSORS

Place the iFeel Sensor so the red light is on top of your fingernail!

 The signal is stronger when the red 

light is on top of your fingernail (there is another, infrared light shining from below). If the sensor doesn’t work 
when placed with the red light on your fingernail can try rotating it so the red light is under your fingertip. 
Make sure the sensor is not too loose or tight on your finger. Try both of your sensors to make sure the 
problem is not the sensor. 

CHECK BLUETOOTH CONNECTION

: Problems with the Bluetooth connection can result in a red heart rate line, 

missing beats, or strange raw Blood Volume Pulse displays. 

Try restarting Alive and restarting the iFeel Sensor.

 If 

restarting doesn’t help, you need to verify your COM settings: 

1)

 

Open Alive with the iFeel Sensor off. Run a graph training session. You will see Waiting for iFeel 
heart rate sensor data on COM [x]. Write down that COM port number. Close Alive. 
 

2)

 

Open Device Manager (click Start, then type "device manager", then click Device Manager). In 
Device Manager click the triangle to expand Ports and find Standard Serial over Bluetooth link. 
You should see the COM port you wrote down in step 1. For example, if you wrote down 5 look for 
"Standard Serial over Bluetooth link (COM5)". 
 

3)

 

Double click on the "Standard Serial over Bluetooth" port, click on the Port Settings tab and 
ensure the settings are Bits per second: 9600, Data bits: 8, Parity: None, Stop bits: 1, Flow control: 
None. You may need to also do this for the other Standard Serial over Bluetooth ports. If you 
change Port Settings please write down the prior values in case you modify port settings for a 
different device and need to revert the changes. 

CHECK BLUETOOTH ADAPTER: 

If you have never had a good signal in Alive on this computer the Bluetooth 

adapter may not be functioning correctly. See the My Computer Doesn’t Have Bluetooth section below. 

 

MY IFEEL SENSOR SHOWS CHARGING EVEN AFTER I REMOVE IT FROM THE CHARGER 

Take the USB charging cable and the iFeel Sensor and walk outside or into another room, away from your 
Bluetooth computers, tablets, or phones. 

With the thin USB side of the charging cable not connected to any charger

 

plug the charging cable into the iFeel Sensor, wait one second, and then remove the charging cable from the 
iFeel Sensor. 

Summary of Contents for Alive iFeel

Page 1: ...iFeel Sensor U S E R G U I D E S U P P L E M E N T ...

Page 2: ... an iFeel Sensor with a rounded square display you have the iFeel Bluetooth Sensor Please start reading from the iFeel Bluetooth Sensor section on the 5th page IFEEL USB SENSOR If you have an iFeel Sensor with a thin rectangular connector you have the iFeel USB Sensor Remove the cap to expose the USB connection Please start reading on the next page to set up your iFeel USB Sensor ...

Page 3: ... attach the small white shirt clip to the blue cable about 10 inches down from the end of the Clip Sensor When reading heart rate from your ear attach the shirt clip to your shirt or jacket to hold the Clip Sensor up so that the Clip Sensor is not pulling down on your ear 3 Optional Attach the long USB extension cord to the iFeel USB Sensor The shorter mini USB converter is for use on Android devi...

Page 4: ...u see a button Change iFeel COM Port click that button and Alive should now find your iFeel Sensor If you do not see this button open the Alive Main Menu click Graph Training from the left side buttons then click Play You should now see the Searching for your iFeel Labs Heart Rate Sensor dialog When you do click Change iFeel COM Port If this still doesn t work and you are using Windows you may nee...

Page 5: ...el Sensor hold the ON OFF button for 5 seconds 4 If the battery is low remove the heart rate sensor and attach the charging cable For quick charging use a 1 Amp or higher wall charger To view the battery indicator remove the charging cable and turn the sensor on 5 A short press on the ON OFF button will bring up the Settings screen On the Settings screen a short press moves to the next setting a l...

Page 6: ...main menu choose Graph Training under Alive workshops and start a session Once a session is running Alive will help you connect the iFeel Sensor to your computer s Bluetooth Follow the steps explained in Alive or follow the Pairing the iFeel Sensor section below 10 For heart rate or smoothness training the Clip sensor can be placed on your finger or ear Please see the iFeel USB Section Clip Sensor...

Page 7: ...ing session Alive will scan for your iFeel Sensor and give you further connection instructions MY COMPUTER DOESN T HAVE BLUETOOTH Somatic Vision does not provide support to help add Bluetooth adapters to computers without Bluetooth Please follow these steps and seek professional assistance if needed 1 Purchase a Bluetooth USB Adapter preferably one with good reviews from a major name brand There a...

Page 8: ...IFEEL SENSOR COM PORT AFTER PAIRING After pairing the iFeel Sensor is assigned two COM Ports These COM Ports should not change during normal use To find and save your COM Port start a graph training orientation or workshop session in Alive Alive will scan COM Ports from 20 down to 3 looking for your iFeel device If Alive can t find the COM Port you will need to manually enter it To manually enter ...

Page 9: ... can result in a red heart rate line missing beats or strange raw Blood Volume Pulse displays Try restarting Alive and restarting the iFeel Sensor If restarting doesn t help you need to verify your COM settings 1 Open Alive with the iFeel Sensor off Run a graph training session You will see Waiting for iFeel heart rate sensor data on COM x Write down that COM port number Close Alive 2 Open Device ...

Page 10: ...stered trademarks of the Institute of HeartMath in the U S and other countries All other trademarks belong to their respective owners Coherence algorithms contained within this product licensed from and 1999 2010 Quantum Intech Inc with all rights reserved and are protected by the following patents in the US 6 358 201 7 117 032 7 163 512 7 462 151 and 8 066 637 Other patents include European Union...