background image

Page 8 of 9 

  

 3118.B 

 

 

 

SPECIFICATIONS 

 

ORS-2000 Series Irrigation Solution Warmer 

 

Power:                                               

 

 

115 VAC; 2A 50-60 H 

                                                                               

     (230 VAC optional) 

 

Fuse:                                                                               2X; 3A Slo-Blo type 

 

Normal Temperature Range:                   Room temperature to 120

F (49

 o

C) 

 

Temperature Variance:                                                               +/-2

o

 F (1

 o

C) 

 

Temperature Control:                         Microprocessor temperature controller; 

                                                                     LED readout of temperature and 

                                                                                      Set Point temperature.  

 

   

    Model 

 

 

Capacity 

 

 

 

Bowl Dimensions   

 
   

ORS-2038         

 

3.8 Liters 

 

     

 

11½" x 5¾" x 4" 

   

ORS-2057 

 

 

5.7 Liters 

 

 

 

19

" x 5¼" x 4" 

   

ORS-2058 

 

 

5.8 Liters 

 

 

 

11½" x 5¾" x 6" 

   

ORS-2066 

 

 

6.6 Liters 

 

 

 

11½" x 9¼" x 4" 

   

ORS-2075 

 

 

7.5 Liters  

(Round Basin) 

14 ½"(O.D.) x 4 ¾"  

           ORS-2094 

 

 

9.4 Liters 

 

 

 

11½" x 9¼" x 6" 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Summary of Contents for ORS-2000 Series

Page 1: ...OPERATORS MANUAL for the ORS 2000 Series Irrigation Solution Warming System O R Solutions Inc 3901 Centerview Drive Suite L Chantilly VA 20151 800 343 6771 703 834 0101 703 834 0966 Fax 3118 B...

Page 2: ...N IF USED IN THE PRESENCE OF FLAMMABLE ANESTHETICS CAUTION FOR THE ORS 2038 AND ORS 2058 WARMING MACHINES DO NOT OPERATE WARMING MACHINE WITHOUT A MINIMUM OF 500 mL SOLUTION IN THE ORS APPROVED DRAPE...

Page 3: ...1 Upper indicator PV Present value indicates present temperature of solution 2 Lower indicator SV Set value indicates the temperature the solution is set to warm to 3 AL lights when Present Value is 4...

Page 4: ...perature to desired setting using up down arrows Step 3 Press enter Step 4 Press index F Depress POWER switch to turn machine off Do not turn off by pulling power cord from outlet G Discard solution a...

Page 5: ...emperature Controller verification The ORS 2000 Series Warmer temperature controller should be verified every 5 years O R Solutions Inc requires the use of the following items or equivalent fluid temp...

Page 6: ...r replace defective warmers Improper use and or treatment of the warmer voids this warranty at the option of O R Solutions Inc O R SOLUTIONS INC DISCLAIMS ALL OTHER WARRANTIES EXPRESSED OR IMPLIED INC...

Page 7: ...ch set point temperature O R Solution Make sure that the temperature is adjusted properly see page 3 If the temperature adjustment is correct inspect the drape The drape should be evenly distributed a...

Page 8: ...120o F 49 o C Temperature Variance 2o F 1 o C Temperature Control Microprocessor temperature controller LED readout of temperature and Set Point temperature Model Capacity Bowl Dimensions ORS 2038 3...

Page 9: ...UNTING INSTRUCTIONS 1 Assembling ORS 2000 Series Solution Warmer with the ORS S21 or ORS S22 Stand 2 Mounting of ORS 2038 Solution Warmer to ORS 1075HS Slush Machine Mounting Screws 2x ORS 2038 Irriga...