background image

Page 3 of 9 

  

 3118.B 

 

 
NOTE: 

GROUNDING  RELIABILITY  CAN  ONLY  BE  ACHIEVED  WHEN 
THIS  EQUIPMENT  IS  CONNECTED  TO  AN  EQUIVALENT 
RECEPTACLE  MARKED  "HOSPITAL  ONLY"  OR  "HOSPITAL 
GRADE." 

 
CAUTION: POWER  MUST  BE  OFF  UNTIL  MINIMUM  FLUID  LEVEL  IS 

POURED INTO DRAPE. 

 

 

 

TEMPERATURE CONTROLS 

 

1. 

Upper  indicator  (PV  -  Present  value)  indicates  present  temperature  of 
solution. 

 
2. 

Lower indicator (SV - Set value) indicates the temperature the solution is 
set to warm to. 

 
3. 

AL lights when Present Value is 4

o

F greater than Set Value. 

 
4. 

o

F indicates solution/set point temperature readouts are in Fahrenheit.* 

 
5. 

SP1 lights when heating element is active. 

 
6. 

SP2 is not applicable. 

 
7. 

 

 used to increase/decrease temperature setting (SV). 

 
*Centigrade readout available upon request. 
 
 
 
 
 
 
 
 

Summary of Contents for ORS-2000 Series

Page 1: ...OPERATORS MANUAL for the ORS 2000 Series Irrigation Solution Warming System O R Solutions Inc 3901 Centerview Drive Suite L Chantilly VA 20151 800 343 6771 703 834 0101 703 834 0966 Fax 3118 B...

Page 2: ...N IF USED IN THE PRESENCE OF FLAMMABLE ANESTHETICS CAUTION FOR THE ORS 2038 AND ORS 2058 WARMING MACHINES DO NOT OPERATE WARMING MACHINE WITHOUT A MINIMUM OF 500 mL SOLUTION IN THE ORS APPROVED DRAPE...

Page 3: ...1 Upper indicator PV Present value indicates present temperature of solution 2 Lower indicator SV Set value indicates the temperature the solution is set to warm to 3 AL lights when Present Value is 4...

Page 4: ...perature to desired setting using up down arrows Step 3 Press enter Step 4 Press index F Depress POWER switch to turn machine off Do not turn off by pulling power cord from outlet G Discard solution a...

Page 5: ...emperature Controller verification The ORS 2000 Series Warmer temperature controller should be verified every 5 years O R Solutions Inc requires the use of the following items or equivalent fluid temp...

Page 6: ...r replace defective warmers Improper use and or treatment of the warmer voids this warranty at the option of O R Solutions Inc O R SOLUTIONS INC DISCLAIMS ALL OTHER WARRANTIES EXPRESSED OR IMPLIED INC...

Page 7: ...ch set point temperature O R Solution Make sure that the temperature is adjusted properly see page 3 If the temperature adjustment is correct inspect the drape The drape should be evenly distributed a...

Page 8: ...120o F 49 o C Temperature Variance 2o F 1 o C Temperature Control Microprocessor temperature controller LED readout of temperature and Set Point temperature Model Capacity Bowl Dimensions ORS 2038 3...

Page 9: ...UNTING INSTRUCTIONS 1 Assembling ORS 2000 Series Solution Warmer with the ORS S21 or ORS S22 Stand 2 Mounting of ORS 2038 Solution Warmer to ORS 1075HS Slush Machine Mounting Screws 2x ORS 2038 Irriga...