Soltron Space Cab - User manual
If the remote control function is enabled, the display will appear as the following page:
1
S o l t r o n
S T A R T
This page will come back after 3 minutes without pressing any button; pressing one of the 4
buttons the following page will appear:
12
T I M E R E M 0 0 : 0 0
S T A R T E S C
press
ESC
for going back to page
1
Pressing START button a free potential switch will be closed (start command), and this switch
will remain closed for 5 seconds and then will open again;
If a start arrives from the external connector, the session begins:
12a T I M E R E M 0 0 : 1 3
Increasing time is displayed
S T O P
At the end of the treatment (or pressing
STOP
button), page
2b
will be displayed (cooling).
REMOTE CONTROL CONNECTION
On the external side you have a 6-pole connector, on the top back of the unit, to allow the interfacing
to the external control.
In the drawing below you can see how to connect the unit to a generic external control.
Soltron Space Cab
230V ac L
N
Input 230Vac L1
For start from N1
external control
Internal
start
1
2
5
6
3
4
External
start
21