SRU User Guide
Page 11
4.0 Real-Time Readings
4.1 Take a Reading
Select
Take a Reading
to instantly display the current water level (water column height above the pressure
sensor), temperature, and conductivity being detected by the connected datalogger.
Note:
Only LTC Leveloggers will display conductivity readings.
Note:
See the Levelogger User Guide (Section 1.2) and Vented Dataloggers User Guide (Section 1.2) for an explanation of the
pressure sensor measurement fundamentals.
Figure 4-1 Take a Reading
The water level reading will indicate "Comp" as it is a compensated level reading. See Section 4.2 for more
details.
4.2 Barometric Compensation
Levelogger 5 dataloggers use an absolute pressure sensor, therefore, detect total pressure above the sensor
(water plus barometric pressure).
To eliminate barometric pressure and display water level only, select Next after taking a reading (Section 4.1).
Select to turn barometric compensation on. The SRU contains an internal barometer that barometrically
compensates the level readings. The barometer reading will be displayed on this screen.
Figure 4-2 Turn Barometric Compensation On/Off
LevelVent 5 and AquaVent 5 vented water level dataloggers use gauged pressure sensors, therefore, barometric
effects are already eliminated from the level reading through the vented cable.
Barometric compensation can be turned off in the same screen.
Next
Back
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle
99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle
99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle
99%
Back
Start
12 mins
Sample Rate:
Compensation: On
Idle
99%
Stop
Save
Level (Comp): 10.045m
Temperature: 28.234
ºC
Conductivity: 125.4
µ
S
Log 56
99%
Home
Data Saved
Compensated
Barometrically
Log 57
99%
Home
Data Saved
Uncompensated
Log 57
99%
Stop
Save
Level (Uncomp): 10.045m
Temperature: 28.234
ºC
Conductivity: 125.4
µ
S
Log 56
99%
Back
Start
12 mins
Sample Rate:
Compensation: Off
Idle
99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading
Download
99%
99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle
99%
Next
Back
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle
99%
Next
Back
LCD Contrast
Logger Status
SRU Status
Idle
99%
Download Logger
Idle
99%
Idle
99%
View Files
Next
Back
Download Logger
View Files
Idle
Next
Back
99%
Next
Back
LCD Contrast
Logger Status
Idle
99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3
Home
3/11 Files
99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
Next
Back
Idle
99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
Next
Back
Idle
99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back
Home
Idle
99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle
99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back
Home
Idle
99%
55%
Contrast:
Set LCD
Next
Back
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle
99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle
99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle
99%
Back
Start
12 mins
Sample Rate:
Compensation: On
Idle
99%
Stop
Save
Level (Comp): 10.045m
Temperature: 28.234
ºC
Conductivity: 125.4
µ
S
Log 56
99%
Home
Data Saved
Compensated
Barometrically
Log 57
99%
Home
Data Saved
Uncompensated
Log 57
99%
Stop
Save
Level (Uncomp): 10.045m
Temperature: 28.234
ºC
Conductivity: 125.4
µ
S
Log 56
99%
Back
Start
12 mins
Sample Rate:
Compensation: Off
Idle
99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading
Download
99%
99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle
99%
Next
Back
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle
99%
Next
Back
LCD Contrast
Logger Status
SRU Status
Idle
99%
Download Logger
Idle
99%
Idle
99%
View Files
Next
Back
Download Logger
View Files
Idle
Next
Back
99%
Next
Back
LCD Contrast
Logger Status
Idle
99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3
Home
3/11 Files
99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
Next
Back
Idle
99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
Next
Back
Idle
99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back
Home
Idle
99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle
99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back
Home
Idle
99%
55%
Contrast:
Set LCD
Next
Back
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle
99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle
99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle
99%
Back
Start
12 mins
Sample Rate:
Compensation: On
Idle
99%
Stop
Save
Level (Comp): 10.045m
Temperature: 28.234
ºC
Conductivity: 125.4
µ
S
Log 56
99%
Home
Data Saved
Compensated
Barometrically
Log 57
99%
Home
Data Saved
Uncompensated
Log 57
99%
Stop
Save
Level (Uncomp): 10.045m
Temperature: 28.234
ºC
Conductivity: 125.4
µ
S
Log 56
99%
Back
Start
12 mins
Sample Rate:
Compensation: Off
Idle
99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading
Download
99%
99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle
99%
Next
Back
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle
99%
Next
Back
LCD Contrast
Logger Status
SRU Status
Idle
99%
Download Logger
Idle
99%
Idle
99%
View Files
Next
Back
Download Logger
View Files
Idle
Next
Back
99%
Next
Back
LCD Contrast
Logger Status
Idle
99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3
Home
3/11 Files
99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
Next
Back
Idle
99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
Next
Back
Idle
99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back
Home
Idle
99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle
99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back
Home
Idle
99%
55%
Contrast:
Set LCD
Next
Back
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle
99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle
99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle
99%
Back
Start
12 mins
Sample Rate:
Compensation: On
Idle
99%
Stop
Save
Level (Comp): 10.045m
Temperature: 28.234
ºC
Conductivity: 125.4
µ
S
Log 56
99%
Home
Data Saved
Compensated
Barometrically
Log 57
99%
Home
Data Saved
Uncompensated
Log 57
99%
Stop
Save
Level (Uncomp): 10.045m
Temperature: 28.234
ºC
Conductivity: 125.4
µ
S
Log 56
99%
Back
Start
12 mins
Sample Rate:
Compensation: Off
Idle
99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading
Download
99%
99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle
99%
Next
Back
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle
99%
Next
Back
LCD Contrast
Logger Status
SRU Status
Idle
99%
Download Logger
Idle
99%
Idle
99%
View Files
Next
Back
Download Logger
View Files
Idle
Next
Back
99%
Next
Back
LCD Contrast
Logger Status
Idle
99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3
Home
3/11 Files
99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
Next
Back
Idle
99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
Next
Back
Idle
99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back
Home
Idle
99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle
99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back
Home
Idle
99%
55%
Contrast:
Set LCD
Summary of Contents for SRU 4001
Page 1: ...Model 4001 Solinst Readout Unit SRU User Guide September 15 2021...
Page 2: ......
Page 4: ......