Levelogger User Guide - Software Version 3.4.0 or Higher
Page 31
Calibration:
Since all standard calibration solutions state conductivity at a
standard temperature of 25˚C, the LTC Levelogger Junior can account for
temperature differences between 15 - 30˚C when you calibrate the unit. The
LTC Levelogger Junior will compare the current temperature and conductivity
readings against the temperature corrected standard solution. When a single-
point calibration is successful the process is completed. If a failure does occur
within the singe-point calibration procedure. The LTC Levelogger Junior will
default back to the last pre-calibrated state. When conducting a multipoint
calibration, once the first calibration point is successful the process repeats itself
to the next calibration point (back to the beginning of Step 3). If a failure does
occur within the multi-point calibration procedure, the LTC Levelogger Junior
will accept all of the successful calibration points and default back to the last pre-
calibrated state for any failed ones.
Figure 6-6 LTC Calibration Wizard Progress
Figure 6-7 LTC Calibration Wizard Results
Note:
Once the
calibration is
complete for the
first solution of a multipoint
calibration, the Calibration
Wizard will then go back to
the beginning of Step 3 to
start the rinsing process for
the next solution selected.