![Sofraser MIVI 8001 Technical Manual Download Page 35](http://html1.mh-extra.com/html/sofraser/mivi-8001/mivi-8001_technical-manual_1312041035.webp)
Technical Manual MIVI 8001
Ref. 128-41/5
35
CNOMO WORDS
Addresses (hexa)
r/w
Observations
WORD Scale
000B r
CNOMO Min Scale (pd = 1 without decimal, pd = 10 one
decimal ... pd = 1000 three decimals)
* PD
000C
r
CNOMO Max Scale
* PD
000D
rw CNOMO Threshold alarm 1
* PD
0079
r
CNOMO Manufacturer brand
Most significant bit :
13d
007A
r
CNOMO Units type
Most significant bit :
20d
0912-0915
rw
!
THRESHOLD : alarms threshold 1 to 4.
If the threshold is declared : limited writing(
*
)
Extent according to the configuration min/max process variable
0922-0931 rw
CONST: Program constants
(-999.0/9999)
0932
rw
∗
Max. for viscosity bargraph
(-999.0/9999)
0933 rw
Max. for temperature bargraph
0942 rw
Viscosity analog output
(Min/Max scale)
0943 rw
Temperature analog output
(Min/Max scale)
0946 r
Minimum for viscosity output scale
(-999/9999)
0947 r
Minimum for temperature output scale
(-999/9999)
094A r
Maximum for viscosity output scale
(-999/9999)
095B r
Maximum for temperature output scale
(-999/9999)
094F rw
Viscosity measurement
(-999/9999)
0950 rw
Temperature measurement
(-999/9999)
0959 rw
Sensor's frequency
(AFFR_/AFFR-)
0963 rw
Minimum frequency = AFFR_
(-999/9999)
096D rw
Maximum frequency = AFFR-
(-999/9999)
098F rw
Global locking.
Automatically goes back to its initial value if the unit has not received
any order.
Entire number
(0-4)
0993 rw
Locking ALARM block (strong weight of the word) and FILTRE
block (weak weight of the word).
Automatically goes back to its initial value if the unit has not received
any order for one minute.
Entire number
(0-2)
0995 rw
Locking PARAM block (strong weight of the word) and CONST
block (weak weight of the word).
Automatically goes back to its initial value if the unit has not
received any order for one minute.
Entire number
(0-2)
0997 rw
Locking M.CARD block (strong weight of the word) and
HORLOG block (weak weight of the word).
Automatically goes back to its initial value if the unit has not
received any order for one minute..
Entire number
(0-2)
The stored values in relative value to the addresses 0800H to 09FFH can be reached in absolute
value (IEEE = 2 registers). You have to apply the conversion formula below so as to determine their
address.
!
According to the configuration, the parameter can be saved in a EEPROM (100000 writtings) or in NOVRAM (unlimited
writings).
∗
Parameter is saved in EEPROM and can only be written 100000 times. The other ones have an unlimited writing times.