It can be zinc-/coal- and alkaline/manganese batteries and nickel-cadmium and nickel-metal-
hydride rechargeable batteries.
In order to change the battery, open the battery compartment on the underside of the baby
scale.
3. FAULTS – CAUSES AND REMEDIES
DISPLAY
CAUSES
REMEDY
Underload: only dashes appear at the bottom
of the display field.
Switch the scale off and on again. If
the error occurs again, please contact
your service partner.
Overload: only dashes appear at the top of
the display field. Maximum weighing range is
exceeded.
Unload the scale, or switch them off
and on again. Check the maximum
weighing range.
The battery is empty.
Replace the battery.
No standstill of scale.
Check the surface and the leveling of
the weighing platform. Check the
scale on a force shunt. If the error
message after the realignment
persists, contact your service
provider.
If the fault cannot be cleared, or other error messages appear, please contact your service
partner.