21
Max. weighing range exceeded
> Remove weighing product from
scale
Set to zero not possible
> Tolerance to zero point too big
■
When calibrating
When the scale was unloaded
during switch on
initialization
> Switch scale off and on again
Calibration weight too low
Tolerance is too high to fixed
weight level (e.g.
5,2 kg instead of 5,0 kg)
5. ERROR MESSAGES
■
User calibration
• Switch scale on and press
while switching on.
• Enter pin 0000 and press
.
• Enter kg unit and press
.
•
Press
when the unloaded
scale is in still stand (arrow sign
points to
).
•
Place the calibration weight on
the scale. Press
when the
scale is in still stand (arrow sign
points to
).
Use a calibration weight with min. 15% of the weighing
range, e.g. for 45 kg weighing range you can use a
calibration weight of 5 – 45 kg.
Calibrate in the range where you can get the highest
accuracy.
• Switch on initialization starts. During this activity
unload the scale.