HI OK LO
R1
R2
R3
0
NET
g
GN
ct
dwt
mm
lb
%
oz
tl
J
H
M
A
SAMPLE
On/ZERO
TARE
MODE
Off
CHARGE
Outline descriptions
1. Turns balance On.
2. Captures a new center of zero.
Reduce gross weight on pan as tare weight.
1. Long press for selecting various units of weight.
1. Short press for parts counting and percentage weighing.
2. Short press to toggle switch for display total weight,
quantity or percentage.
Print out.
Function keys
Weighing Pan
Leveling Adjust
Adaptor Port
AC
Numeric Display
Function
SAMPLE
On/ZERO
TARE
MODE
Off
Turns balance Off.
3
8
Calibration
The window display:
2. Long press the key around 5 seconds.
The calibration process will run step by step automatically.
a._ The setting weight value appears.
b._ Put on the Calibration weights on the pan.
1. Keep empty on the pan.
3. If the different weight is used for this calibration,
The displayed weight should be within the range of Calibration w/- 0.05% of Max.
The error message appears and the balance
4. Check the weight and retry the weight calibration.
Then weight calibration can be work.
g._ The weight calibration to be done.
e._ Unloading Calibration weights on the pan.
d._ Span calibrate done and go Zero calibration.
c._ Span calibrating.
f._ Zero calibrating.
returns to weight display.
0
g
0
g
0
g
E. Print out
Manual
Auto
F. Baud rate
2400
9600
Off
G. Cal. Weight setting
Calibrate done
Programme
200 g
1200 g
.
.
.
.
2. Long press for comparison mode setting.