7.0 USER CALIBRATION
The EEWB333B Balancer features a User Calibration
program which requires only a few minutes to com-
plete. The operator can electronically calibrate the
rotating masses on the machine; which is called User
Calibration.
Perform this procedure when the balancer has been
moved, disturbed, or whenever accuracy is questioned.
If the calibration process fails a service technician must
be called to perform a complete calibration.
A calibration run requires little more time than a regu-
lar balance run.
Procedure
;
—
Balance a wheel, as a NORMAL weight mode, to
less than 5 grams per plane.
—
Press the “
MODE
” key and the “
FINE
” key together
for 5-7 seconds. See Figure 30a.
The initial display reads “
C 14
”, See Figure 30b.
—
Press “
ENTER
” once to initiate procedure.
The display shows “
CAL 1
”. See Figure 30c.
—
Remove any cones or adapters from the shaft. See
Figure 30d.
Note:
Continue the spin until the unit beeps.
When compete, the display shows “
CAL 2
”. See Figure 30e.
—
Mount the User Calibration Weight, refer to Figure
30f. Spin the shaft.
—
After a few seconds the display equals
C ---
, Figure 30g.
The User Calibration was performed correctly.
Any (operator) error causes the program to exit. Un-
screw the Calibration weight from the
fl ange and put it
back in its designated place for safe keeping.