NOTE:
Chamber pressure must be released before the pressure chamber door can be
opened.
To Clean
Please refer to the maintenance section of this manual for complete cleaning and
disinfecting procedures for the H-1025 pressure infusers.
Disposables: Set up and Performance
! WARNING !
Refer to D-series disposable packaging for complete set up and priming instructions,
indications, contraindications, warnings, cautions and additional product information.
Setting Up Disposable Sets
Level 1
®
D-series disposable sets have been designed to be highly effective and easy to
use in combination with the H-1000 fluid warming hardware. By following the simple 1-2-3
blocks found on the H-1000 fluid warmer, properly trained users can confidently have the
system safely set up and operating within seconds.
Following the numbers found on the H-1000:
1. Push the bottom end of the heat exchanger into the bottom heat exchanger #1 socket.
Be sure the heat exchanger is fully seated.
! WARNING !
Do not bend heat exchanger. Bending the heat exchanger may damage the inner
aluminum tube. If the tube is bent, serious patient injury or death could result.
2. Slide the top (#2) socket up. Snap the heat exchanger into the heat exchanger guide.
Be sure it is fully seated. Slide the top heat exchanger socket down over the heat
exchanger tube.
3. Insert the Filter with Gas Vent into the #3 holder on the front of the cabinet.
4. Press the green power switch on the front of the cabinet to turn ON the unit and allow it
to reach operating temperature while the disposable set is being primed. It takes
between 3 and 7 minutes for the water in the H-1000 to reach operating temperature.
12