Chapter 9: Troubleshooting
9-2
Home Use Operation Manual
Power
PRObleM
POSSIble CAUSe
CORReCTIVe ACTIOn
The oximeter doesn’t
turn on.
Batteries weak and/ or no AC
power.
Batteries not installed or batteries
incorrectly installed.
Replace or recharge the batteries.
Connect to source of AC power.
Ensure the batteries are installed
correctly. Connect to source of AC
power.
The oximeter turns
on, but display reads:
L0
Bat
AC power is connected, but no or
weak batteries are installed.
This is proper operation.
Replace or recharge the batteries.
The oximeter turns
off unexpectedly.
Batteries are weak or dead and no
AC power source is connected.
Replace or recharge the batteries.
Connect to source of AC power.
Cannot turn off
monitor with ON/
OFF key.
Monitor is in Home “
K
” mode.
This is proper operation. To turn the
monitor off while in Home Mode,
press and hold the Exit key and
then press the ON/OFF key.
The External Power
indicator
X
is not lit.
(AC Power).
An External AC Power Supply cord
is not connected to the monitor
or accessory cable. The accessory
cable is not connected to the
monitor.
The AC power cord is connected
to a wall outlet that is controlled
by a wall switch.
A non user-serviceable fuse may
have opened.
Check all power supplies and cables
for proper connections.
Only connect the AC power cord to
an outlet that is not controlled by a
wall switch.
Contact your authorized service
representative.
The External Power
indicator
X
is not lit.
(USB Cable)
The USB cable is not connected
to a source of power or the USB
power source is not operating.
Check all cable connections and
confirm PC, USB hub or other
source of USB power is working
correctly.
The USB Power
indicator
is not
lit. (Docking Station)
AC Power is connected.
The USB cable is not connected
to a source of power or the USB
power source is not operating.
This is proper operation.
Check all cable connections and
confirm PC, USB hub or other
source of USB power is working
correctly.
The monitor
operates on AC
power, but not on
battery power.
The battery is missing.
The battery is drained.
The battery is defective.
Replace the battery.
Charge or replace the battery.
Contact your authorized service
representative if using LI-Ion
rechargeable battery pack.
Summary of Contents for Oxi-Pulse 30
Page 2: ......
Page 6: ...Table of Contents iv Home Use Operation Manual This page is intentionally left blank ...
Page 16: ...Chapter 1 Introduction 1 10 Home Use Operation Manual This page is intentionally left blank ...
Page 22: ...Chapter 3 Getting Started 3 4 Home Use Operation Manual This page is intentionally left blank ...
Page 46: ...Chapter 6 Alarms 6 6 Home Use Operation Manual This page is intentionally left blank ...
Page 62: ...Chapter 9 Troubleshooting 9 6 Home Use Operation Manual This page is intentionally left blank ...
Page 65: ......