Programming and Operation
Yes
•
No
Programming and Operation
Security Settings
The security settings are used to limit patient and clinician access to certain programming and operating
functions of the pump. The various functions of the pump are protected by three different security codes,
and may also be protected by the cassette/keypad lock. The security level table below lists the functions that
are available under each security code. The factory default settings for the security codes are as follows:
*
*
Text omitted from online version **
Administrator Code
: See your CADD
®
-Solis system administrator.
The keypad code, clinician code, and administrator code can all be customized by the CADD
®
-Solis system
administrator while setting up the protocol using the CADD
®
-Solis Medication Safety Software, or in the
administrator settings. The CADD
®
-Solis system administrator also determines whether or not to allow
use of the cassette/keypad lock to unlock the keypad. See your CADD
®
-Solis system administrator for more
information or to learn the security code you should use if the codes have been customized.
Security Level Table
The keypad code should be used by clinicians who need to set up and manage a protocol for a patient. The
clinician code will allow access to all the functions the keypad code allows, as well as the clinician bolus
feature. The administrator code will allow access to all functions of the pump and its use should be restricted
to the CADD
®
-Solis system administrator and certain designees. The administrator code gives the user the
ability to change protocol ranges and all settings in the pump.
Pump Operations
and Programming
Keypad Security Locked
Keypad Security
Unlocked
Clinician Security
Unlocked
Administrator
Security Unlocked
Running
Stopped
Running
Stopped
Running
Stopped
Running
Stopped
Stop/Start
•
•
•
•
•
•
•
•
Reset Reservoir Volume
•
•
•
Start PCA Dose
•
•
•
•
Change (Titrate) Continuous Rate
•
•
•
•
•
•
Change (Titrate) PCA Dose
•
•
•
•
•
•
Change (Titrate)
•
•
•
•
•
•
Change (Titrate) Delivery Amount
•
•
•
•
•
•
Change (Titrate) Max Doses/Hr
•
•
•
•
•
•
Pump Tasks
Keypad Security Locked
Keypad Security
Unlocked
Clinician Security
Unlocked
Administrator
Security Unlocked
Running
Stopped
Running
Stopped
Running
Stopped
Running
Stopped
Give Clinician Bolus
•
•
Start New Patient
•
•
•
Start New Protocol/Same Patient
•
•
•
Prime Tubing
•
•
•
Set Time and Date
•
•
•
Adjust Backlight Intensity
•
•
•
•
•
•
Adjust Alarm Volume
•
•
•
•
•
•
View/Clear Reports
•
•
•
•
•
•
•
•
Access Administrator Settings
•
Security Level Table Key