18
Note: The communication module may be interfered
with by other equipment, even if that other
equipment complies with CISPR 11 requirements.
For additional CADD®-Solis Communication Module
network specifications and setup information, see
the Network Setup Utility Installation and User
Guide (Version 1.1).
Interference / Separation Distances
Other electronic communication devices used in
the vicinity of the communication module may
cause interrupted transmission of data. Interrupted
transmission of data does not impact pump delivery.
The table below specifies what effect each type of
device causes and describes the minimum separation
distance that should be observed.
Interferer (type of electronic
device)
Min.
Separation
Distance
2.4GHz/5GHz Wireless Transmitter
Base Stations
(e.g. Wi-Fi® Access
Points, Wireless Telephone Base
Stations)
3 meters
(9.84 feet)
2.4GHz/5GHz Wireless Medical
Devices and other Peripherals
(e.g.,
Wi-Fi® enabled patient monitors,
infusion pumps, computers/smart-
phones/tablets; Bluetooth-enabled
devices; Zigbee-enabled devices;
Wireless telephone handsets)
1 meter
(3.28 feet)
Other Smiths Medical CADD®-Solis
pumps with Wireless Connectivity
0.3 meters
(1 foot)
433MHz/2.4GHz/5GHz RFID Asset
Location Tags
Direct
contact
Communication Module Wireless Range:
36.58 meters (120 feet) (line-of-sight operation
with access point)
Network Modes and Encryption
Network Modes:
Encryption Modes:
• 802.11 a/b/g/n
• None
• 802.11 b/g/n
• TKIP
• 802.11 a/n
• CCMP
• TKIP/CCMP
Authentication Modes: EAP Modes:
• OPEN
• None
• WPA-ENT
• EAP-TLS (v1.0 and v1.2)
• WPA2-ENT
• EAP-TTLS
• WPA-PSK
• EAP-PEAPV0 (PEAP with
MS-CHAPv2)
• WPA2-PSK
• EAP-PEAPV1 (PEAP with
GTC)
• EAP-LEAP
• EAP-FAST
Certificates:
SSL SHA-1
SSL SHA-2