Device Operating Modes  

There are three (3) operational modes: Clinician (CLn), Home (H) and Sleep (SLP).

Operating Modes

To enter the mode menu, press and hold the 

1. 

Menu

 key (

 ) at power up as the monitor goes through its 

initialization sequence. CLn (Clinician Mode) will appear.
To change the device’s operating mode, use the up and down keys to toggle between Clinician Mode (

2. 

CLn), Home 

Mode (H) and Sleep Mode (SLP).

trend Interval

To change the Trend Interval, the time interval at which real time data is sent to the device’s memory, enter the 

1. 

Operating Mode menu as explained above.
When 

2. 

CLn appears on the display, press the 

Menu

 key (

 ) again. SeC will appear.

Use the up and down keys (

3. 

n

 

o

) to change the trend interval (2-30 seconds).

To exit, use 

4. 

Exit

 key ( 

 ) or wait 20 seconds.

Language

To change language, refer to WW1924 Operator’s Manual, Section 5.

Remote Alarm

To activate or deactivate the Remote Alarm function, enter the 

1. 

Operating Modes

 menu.

Press and hold the 

2. 

Menu

 key (

 ), then advance three more times. The display will show rA.

Use the up and down keys (

3. 

n

 

o

) to select On or Off.

To exit use the 

4. 

Exit

 key ( 

 ) or press the 

Menu

 key (

 ) until the software version number appears. The oximeter 

will then initialize normally.

Factory Defaults

To restore factory default settings, enter the Operating Modes menu.

1. 

Press the 

2. 

Menu

 key (

 ) four more times. The display will show Fd. 

Use the up and down keys (

3. 

n

 

o

) to select Yes, which restore the factory default settings.

Clearing trend Memory

To clear the oximeter’s memory, turn the unit off.

1. 

Press and hold the 

2. 

On/Off

 button ( 

x

 

) until CLr appears and stops flashing.

If the 

3. 

On/Off

 button ( 

x

 

) is released before CLr stops flashing, memory will not be cleared.

Page 3 of 4

Quick Reference Guide

 

WW1030

Summary of Contents for BCI SPECTRO2 30

Page 1: ... alarm tone briefly press the Alarm Silence key 2 B The Alarm Silence indicator will light solid for the first minute and 45 seconds then flash the last 15 seconds 3 To cancel the 2 minute silence briefly press the Alarm Silence key 4 B Note The Alarm Silence key will not affect visual alarm indicators or pulse tone Changing the Alarm Limits To change the alarm limits press the 1 Menu key until th...

Page 2: ... up and down keys 2 n o to select the year Range is 00 2000 to 99 2099 Press the 3 Menu key to advance to the Month Menu The display will show dt 2 Date Setting 2 Use the up and down keys 4 n o to select the month Range is 01 January to 12 December Press the 5 Menu key to advance to the Day Menu The display will show dt 3 Date Setting 3 Use the up and down keys 6 n o to select the day Range is fro...

Page 3: ...it 20 seconds Language To change language refer to WW1924 Operator s Manual Section 5 Remote Alarm To activate or deactivate the Remote Alarm function enter the 1 Operating Modes menu Press and hold the 2 Menu key then advance three more times The display will show rA Use the up and down keys 3 n o to select On or Off To exit use the 4 Exit key or press the Menu key until the software version numb...

Page 4: ...ring and 2 gently push the positive end until it contacts the positive tab If using the rechargeable lithium ion battery pack WW1090 align the battery pack so that its metal connectors 3 line up with the corresponding connectors in the battery compartment Push battery pack into place Replace battery door by sliding it back into place until the latch clicks 4 Attaching the Sensor Cradle The WW1030 ...

Reviews: