Device Operating Modes
There are three (3) operational modes: Clinician (CLn), Home (H) and Sleep (SLP).
Operating Modes
To enter the mode menu, press and hold the
1.
Menu
key (
) at power up as the monitor goes through its
initialization sequence. CLn (Clinician Mode) will appear.
To change the device’s operating mode, use the up and down keys to toggle between Clinician Mode (
2.
CLn), Home
Mode (H) and Sleep Mode (SLP).
trend Interval
To change the Trend Interval, the time interval at which real time data is sent to the device’s memory, enter the
1.
Operating Mode menu as explained above.
When
2.
CLn appears on the display, press the
Menu
key (
) again. SeC will appear.
Use the up and down keys (
3.
n
o
) to change the trend interval (2-30 seconds).
To exit, use
4.
Exit
key (
) or wait 20 seconds.
Language
To change language, refer to WW1924 Operator’s Manual, Section 5.
Remote Alarm
To activate or deactivate the Remote Alarm function, enter the
1.
Operating Modes
menu.
Press and hold the
2.
Menu
key (
), then advance three more times. The display will show rA.
Use the up and down keys (
3.
n
o
) to select On or Off.
To exit use the
4.
Exit
key (
) or press the
Menu
key (
) until the software version number appears. The oximeter
will then initialize normally.
Factory Defaults
To restore factory default settings, enter the Operating Modes menu.
1.
Press the
2.
Menu
key (
) four more times. The display will show Fd.
Use the up and down keys (
3.
n
o
) to select Yes, which restore the factory default settings.
Clearing trend Memory
To clear the oximeter’s memory, turn the unit off.
1.
Press and hold the
2.
On/Off
button (
x
) until CLr appears and stops flashing.
If the
3.
On/Off
button (
x
) is released before CLr stops flashing, memory will not be cleared.