Chapter 11: Specifications
11-8
Clinician’s Operation Manual
Dock Dimensions
DIMENSION
WITHOUT PRINTER
WITH PRINTER
Width:
107 mm (4.2 inches)
160 mm (6.3 inches)
Height:
84 mm (3.3 inches)
84 mm (3.3 inches)
Depth:
109 mm (4.3 inches)
109 mm (4.3 inches)
Weight:
570 grams (20 ounces)
850 grams (30 ounces)
Auxiliary Inputs/Outputs
Sensor:
DB9
Data or
Power Input
Connector:
Non-standard 16 pin docking connector for power and data.
Environmental
Operating Temperature
Oximeter and accessories: 0°C to 55°C (32°F to 131°F)
Printer: 0°C to 50°C (32°F to 122°F)
Power Supply: 0°C to 40°C (32°F to 104°F)
Li+ battery charging: 5°C to 45°C (41°F to 113°F)
Operating Humidity
Oximeter and accessories: 15 to 95% (non-condensing)
Printer: 20 to 85% (non-condensing)
Power Supply: 5 to 95% (non-condensing)
Storage Temperature
Oximeter and accessories: -40°C to +75°C (-40°F to +167°F)
WW1090 LI-Ion rechargeable
battery pack: -20°C to +60°C (-4°F to 140°F)
Printer: -25°C to +70°C (-13°F to +158°F)
Power Supply: -10°C to +70°C (14°F to 158°F)
Storage Humidity
Oximeter and accessories: 10 to 95% (non-condensing)
Printer: 10 to 90% (non-condensing)
Power Supply: 5 to 95% (non-condensing)
Altitude:
3050 m (10,000 ft) max
Shock and vibration:
ISO 9919 transport rated
NOTE! The monitor may not meet performance specifications when stored or used outside
the temperature and humidity ranges listed above.
Summary of Contents for BCI SPECTRO2 10
Page 3: ......
Page 7: ...Table of Contents iv Clinician s Operation Manual This page is intentionally left blank ...
Page 82: ......