S E C T I O N 1 0
•
M a i n t e n a n c e
34
HOTLINE
®
3 Blood and Fluid Warmer | Operator’s Manual
Maintenance Log
All maintenance and testing should be done by competent personnel.
Regularly scheduled maintenance ensures proper functioning of the
equipment. Refer to the table below for required tasks and frequency of
routine maintenance.
Maintenance Checklist
Task
Every Use Every 30 Days Every 12 Months
Clean the Exterior
General Inspection
Disinfect the Reservoir and
Change Distilled Water or
Isopropyl Alcohol solution
Disinfect the Reservoir and
Change the Hydrogen
Peroxide Solution
Alarm Signal Test
Add Recirculating Solution
Test
Check Disposables Test
Over Temperature Alarm
Test
Verify Temperature
Calibration
Electrical Safety Tests