22
5.9.3
PELVIC LAP STRAP
The
position
and
length
of the pelvic lap strap can be adjusted.
5.9.3.1
POSITION
To alter the position of the CAM locks that hold the lap strap:
Locate the CAM locks (26 & 27) underneath the chair.
Pull the top of the CAM lock so that the head of the fixing screw is exposed.
Hold the nyloc nut that is behind on of the CAM locks (26 & 27) using the supplied
10mm spanner.
Loosen the screw, by turning it anti-clockwise, using the supplied S-key.
Slide the CAM lock to the desired position.
Use the supplied S-key and spanner to tighten the fixing screw which will lock the
CAM lock in position.
Push the top of the CAM lock back down.
Repeat for the other CAM lock.
5.9.3.2
LENGTH
Push the strap ends through the slots in the seat base until they are clearly visible
underneath the chair.
Pull the tops of the CAM locks (26 & 27) so that is perpendicular to the base.
Feed the ends of the straps through the CAM Locks (26 & 27) so that they are at the
desired length.
Push the tops of the CAM locks (26 & 27) so that they are closed.
26
27