7596040 – MANUALE WO-20 EN 21052018 REV.1
- 10 -
3.2
EXCHANGE OF DRAIN CAPILLARY
550
150
490
1
3
5
2
4
6
7
1
2
3
1
2
WARNING
After replacing the capillary, set the filter change time in the control unit (see § 4.8)
WARNING
BLUE CAPILLARY = recovery 25-30% a 15°C
BLACK CAPILLARY = recovery 20-25% a 15°C
WHITE CAPILLARY = recovery 15-20% a 15°C
The equipment must not be installed with inlet water> 1200 mS or> 35 ° dH.
Maximum recovery rates refer to incoming water at 25 ° F.
pH
8.5-9.5
BLACK
55h
WHITE
50h
SOFTENER NEEDED
7.5-8.5
BLUE
60h
WHITE
50h
60h
6.5-7.5
BLACK
55h
WHITE
50h
0-6
(°F)
6-15
(°F)
15-20
(°F)
20-25
(°F)
WARNING
Using a small screwdriver, remove the fixing o-rings (1) from clip(2) of fitting.
Push the fitting clip and remove the white capillary.
Insert the o-rings into the capillary to be installed in the machine.
Then insert the ends of the tube into the fittings.
Using a small screwdriver, put the fixing o-rings below the fitting clip.