SMC Networks CBW1026 General Specifications Download Page 3

SMC Model 1026 Engineering Specification           

 

Rev. 100812              

 

 

Page 3 of 12                                                                                                                               

 
6.3 

All  outputs  external  to  the  electrical  box  shall  be  low  voltage  DC,  Intrinsically  Safe  by 

the definition of OSHA and National Electrical Code (NEC). 
 
6.4 

All  devices  and  components  that  consume  electrical  power  shall  be  UL  listed  and/or 

approved,  and  shall  bear  NEMA  (US  Standards),  IEC  (International  Standards)  or  other 
recognized International ratings appropriate for the use intended.  The control assembly shall be 
produced in a UL Panel Shop and shall be UL listed. 
 
6.5 

Fault Conditions shall be annunciated on the operator interface screen, and shall be sent 

to the RS232 interface port or the Ethernet port for printing and/or data acquisition. At least five 
(5) fault conditions shall be so identified. 
 
 

6.5.1  Emergency  Stop  -  Depression  of  the  Emergency  Stop  Button  shall  reset  the 
machine  and  display  a  fault  message  that  identifies  the  source  of  the  emergency  stop 
signal. 

 
 

6.5.2  Door Open - Lack of positive door closure shall cause the machine to enter Pause 
mode, from where the current cycle can be recovered. Fault message shall indicate which 
door is at fault, if the machine is so equipped. 

 
 

6.5.3  Drive  Blocked  -  Failure  of  the  drive  to  traverse  the  cabinet  in  the  time  allowed 
shall  cause  the  machine  to  enter  Pause  mode  and  shall  cause  the  screen  to  display  the 
Drive Blocked message. 

 
 

6.5.4  Sump Heat - Failure of the sump to reach the desired temperature in the expected 
time shall cause the machine to enter pause mode and shall cause the screen to display the 
Sump Heat Fault message. 

 
 

6.5.5  Fill Fault - Fill Fault message shall be displayed and Pause mode shall be entered 
whenever the sump fails to fill in the allotted time.     

 

7.0 

Pneumatically Operated Valves

  

 
7.1 

All automatic ball valves shall be controlled by pneumatic operators. Electric-motor-

operated ball valves shall not be acceptable. 
 
7.2 

All automatic water and steam control valves shall be direct-operated pneumatic types. 

Pilot-operated solenoid valves or any valve that incorporates a diaphragm-type sealing system 
shall not be acceptable. 

Summary of Contents for CBW1026

Page 1: ...ing as well as many features unique to our design 2 0 Dimensions W X H X L W X H X L MODEL DOOR CONFIGURATION CHAMBER SIZE EXTERNAL OVERALL SIZE 1026 VERTICAL LIFT 48 X 32 X 34 72 X 78 X 38 1 2 1026L...

Page 2: ...viously entrained debris shall not be acceptable 4 5 Strainer shall be easily removable with no tools via a sanitary style clamp System shall be completely self regulating and require routine inspecti...

Page 3: ...al 6 5 2 Door Open Lack of positive door closure shall cause the machine to enter Pause mode from where the current cycle can be recovered Fault message shall indicate which door is at fault if the ma...

Page 4: ...shall be programmable from the Color Touch Screen PIN screens shall be available at the discretion of the supervisor to lock out access to the cycle phase programming functions All timers in the prog...

Page 5: ...he last rinse shall be used to remove heavy soil flushing any easily removed matter to drain in order to ensure that the detergent solution stays as clean as possible Treatment shall be under pressure...

Page 6: ...engineered part and sub assembly shall be scrutinized and all specific design decisions shall be made in the light of these basic criteria 30 1 3 All purchased components shall be un modified off of...

Page 7: ...ets shall be elliptical spray jets equal to Spraying Systems Co 1 4USS8030 No proprietary jets shall be acceptable 30 10 Washer programmable control shall provided by a modular industrial type program...

Page 8: ...provided with a door at each end for operation within a clean dirty room environment 40 5 Door Interlocks Doors shall be provided with pneumatically operated door locks arranged such that the two door...

Page 9: ...de for Direct Factory Accessible Programming This option shall allow trained SMC technicians to directly access the programming of the machine from a remote location for changes or diagnostic purposes...

Page 10: ...be chosen when available heights will not accommodate the standard doors in the open position 40 22 Load Shelf Type Door s Load Shelf Type door s shall be provided in order to match owner s existing...

Page 11: ...ry shall be consulted for steam pressures below 30 psi dynamic 60 1 4 Hot water temperatures of less than 180 degrees F may impact treatment cycle times Factory shall be consulted for recommendations...

Page 12: ...BW1026 1026LS 1026C CBW1026L 1026LLS 1026LC Shipping Wgt 2100 2400 Dynamic Wgt as Installed 1800 2000 Cage Capacity Std Mouse 32 48 Std Rat 12 16 Sump Capacity 25 Gals 25 Gals Heat Radiation typ door...