48
Measuring principle
The
smartLAB
®
sprint nG
Self-Monitoring Blood Glucose System is
designed to allow rapid measurement of blood glucose by using an
electrochemical biosensor technology. This system employs a dis-
posable dry reagent strip technology, based on the FAD-binding
glucose dehydrogenase.
Each test strip features an electrode containing FAD-binding glu-
cose dehydrogenase. A capillary blood sample is applied to the col-
lecting area of the strip, and is automatically drawn into the reac-
tion zone, where the FAD-binding glucose dehydrogenase catalyzes
the glucose dehydrogenation. It will produce gluconolactone.
During the reaction, a mediator transfers electrons to the electrode
surface and generates a current. The amount of the current is pro-
portional to the amount of glucose present in the blood sample.
After 5 seconds, the
smartLAB
®
sprint nG
will show glucose concen-
tration on the screen.
Intended use
The
smartLAB
®
sprint nG
is intended for the quantitative measure-
ment of glucose in venous whole blood or fresh capillary whole
blood from fingertips, palm and forearm. Testing is done outside
the body (In citro diagnostic use). It is indicated for self-testing
(over the counter [OTC]) by persons with diabetes, or in clinical
settings by healthcare professionals, as an aid to monitor the ef-
fectiveness of diabetes control. The meter should be used only with
smartLAB
®
nG
blood glucose test strips.