background image

AIR

AIR

Discard of the cleaning powder 

( Refer to the Operation Manual 7-1~7-2 )

• Cleaning of smartGun PERIO after use and regular maintenance are required.

( Failure to do so may cause powder blockage, water leakage, etc. )

• The Nozzle Tip is a single use product. Do not reuse or resterilize it.
• If powder remains in the Powder Case or Handpiece, it will solidify which may cause clogging.
• Before use, thoroughly dry the coupling, handpiece, hose joint section of the powder case, inside of the powder case, and 

nozzle. Also, perform idle running with air alone to expel any internal water.

• Make sure you clean the smartGun PERIO in an ultrasonic bath once a week to prevent the instrument from clogging ( refer to 

the Operation Manual 8-1 ).

Caution

1

2

After removing the 
cleaning powder in 
the powder case, 
perform idle running 
with air alone to 
discharge water 
and powder in the 
nozzle.

Blow off the powder 
remaining in the 
powder case with air. 

Fit the supplied 
blower nozzle to 
the coupling and 
attach to the 
handpiece. Run idle 
with no water for 
more than 5 sec. to 
clear lines.

Use the blower 
nozzle in the same 
way to clear the lines 
in the smartGun 
PERIO nozzle. Blow 
from the far end of 
the nozzle.

1

3

2

Ultrasonic cleaning

( Refer to the Operation Manual 8-1 )

Be sure to do this once a week or after operating Troubleshooting.

Sterilization

( Refer to the Operation Manual 8-2 )

Dry the O-ring section 
of the powder case 
and the Adjustment 
Ring section, apply a 
drop of the supplied 
turbine oil and rub it 
in with your fingers
( refer to the Operation 
Manual 7-4 ).

Carefully clean the 
inside of the joint 
section of the 
handpiece and the 
O-ring section of 
the powder case 
with the supplied 
cleaning brush.

If powder adheres 
to the O-ring 
section, remove the 
O-rings, and 
carefully clean the  
O-rings and 
grooves.

Remove the rubber 
packing and 
disassemble the 
powder case cap.

Rinse under 
running water.

Perform ultrasonic 
cleaning with hot 
water ( refer to 
the Operation 
Manual 8-1).

Disassemble the powder case cap and rinse under 
running water.
All the parts are autoclavable. Perform autoclave 
sterilization up to 135˚C after cleaning. 
ex.) Autoclave for 20 min. at 121˚C, or 15 min. at 132˚C, 

or 3 min. at 134˚C.

 In case the sterilizer chamber temperature may exceed 

135˚C during the drying cycle, skip the drying cycle.

Blower nozzles for other brands couplings
are also available.

If you fail to lubricate the O-ring section, the powder case and 
handpiece may no longer swivel freely and water may leak.

Be sure to remove the cleaning powder and clean it prior to ultrasonic cleaning.

Rubber packing

Care After Use

Handpiece

Blower nozzle

Blower nozzle

Nozzle

Coupling

When idling, hold the 
Nozzle and Handpiece 
tight.

1

Cleaning

( Refer to the Operation Manual 7-3~7-4 )

2

3

3

4

AIR