Scale Calibration
WARNING:
Scale calibration must not be performed unless it is absolutely necesssary. If
scale is no longer measuring accurately, please call technical service.
1.
Storage tank must be empty.
2.
Go to main menu.
3.
Press ESC for 5 seconds (until service menu shows up).
4.
Select SCALE CALIBRATION and press ENTER.
5.
Place an empty storage tank on the scale and press ENTER.
6.
Wait.
7.
LCD will display PLACE TEST WEIGHT.
8.
Place test weight on the storage tank.
9.
If necessary, use UP & DOWN buttons to change the value to match with the weight
used (default is 2000 grams).
10.
Press ENTER.
11.
Wait.
13.
LCD will display COMPLETED.