Control Solution Range mg/dL
Level 2: 93-125
Level 4: N/A
For use with iGlucose glucose meter
• For testing glucose in fresh
capillary whole blood
• For self testing, For single patient use.
• For in vitro testing
• Read instructions before use
• Do not reuse
• Re-cap immediately after use
• Do not store strips outside the vial
• Store between 4C and 30C and
relative humidity of 10-90%
iGlucose® Blood Glucose Monitoring System in the Control solution
mode. If the test result is within the Control Solution Range printed
on the test strip vial label, the iGlucose Monitoring System passes
the quality control test and your iGlucose System is working properly.
Control Solution Range:
Example of Control Solution Range printed on your test strip vial label.
when should a quality
control test be performed?
• To ensure that your iGlucose Blood Glucose Meter and iGlucose
Blood Glucose Test Strip are working properly.
• To confirm that you are following the correct testing procedures.
• To prepare for your initial blood glucose test.
• To check the iGlucose Test Strip or when opening a new vial of
test strips.
• To check your iGlucose Meter after it has been dropped, damaged
or exposed to liquids.
• If you suspect that your test results are inaccurate, or if your test
results are not consistent with the way you feel.
• To practice glucose testing.