Dr.MUSIC 3s (Model: DM-VME03S)
SMD-IFUE-DMV-001 Rev 0
42
4.5.
EMC(Electro-Magnetic Compatibility)
Guidance and Manufacturer ’s Declaration
- Electromagnetic Emissions
–
This
device is intended for use in the electromagnetic environment specified below.
The user of this device should make sure it is used in such an environment.
Emissions Test
Compliance Electromagnetic Environment- Guidance
RF emissions CISPR 11
Group 1
The device uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The device is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/Flicker
emissions
IEC 61000-3-3
Complies
Guidance and Manufacturer’s Declaration
- Electromagnetic Immunity
–
This device is intended for
use in the electromagnetic environment specified below. The user of this device should make sure
it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic
Environment- Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output
lines
± 2 kV for
supply
mains
± 1 kV for
input/output
lines
Mains power quality should
be that of hospital
environment.
Surge
IEC 61000-4-5
±
1 kV differential
mode
±
2 kV common mode
±
1 kV differential
mode
±
2 kV for
common mode
Mains power quality should
be that of hospital
environment.
Voltage dips,
short
<5 % U
T
(>95 % dip in U
T
)
<5 % U
T
(>95 % dip in U
T
)
Mains power quality should
be that of hospital