Dr.MUSIC 3s (Model: DM-VME03S)
SMD-IFUE-DMV-001 Rev 0
35
4. Maintenance and Information of Dr.MUSIC 3s
4.1.
Specification of Dr.MUSIC 3s
Type of Protection
against electric shock
Internally Powered Rating (100 mAh)
Degree of protection
against applied parts
TYPE BF APPLIED PARTS
Internally Powered Rating
3.7Vd.c.(Li-Polymer), 400 mAh
Charging Voltage
DC 5V, 1A
Applied part
Electrode
Input Device
Power/Mode Button, Intensity Button
Operation time
Using the device alone :
20min (FIXED)
Using the device with
Mobile APP :
5~30min
(Default: 20min)
Output voltage
The output stage is from 0 to 35.
Whenever increases the output intensity by one, The
output voltage is increased 1Vpk.
Starting from 4Vpk at Stage 1, finally maximum output
voltage is 38Vpk at 35 stages.
* Rated load resistance: 500
Ω
Dimension
56(W)
㎜×56
(D)
㎜×12.8
(H)
㎜
Length of cable : 200mm
Electrode(BWX4500, YXN45AX) :
Ø45mm
Equipment Weight
54.5g
Transportation and Storage
environment
Temperature range
-25 ~ 70
℃
Relative humidity
10~93%R.H. (non-condensing)
Atmospheric pressure
700 ~ 1060 hPa
Operating environment
Temperature range
5
℃
~ 40
℃
Relative humidity
15~93% R.H. (non-condensing)
Atmospheric pressure
700 ~ 1060 hPa
Classification
-
For continuous operation.
- Not suitable for use in the presence of flammable
anesthetics.
- Not for use in the presence of an oxygen-enriched
atmosphere (oxygen tent).