![Slendertone Pulse 530 Instruction Manual Download Page 19](http://html1.mh-extra.com/html/slendertone/pulse-530/pulse-530_instruction-manual_1293019019.webp)
FCC/IC transmitter declaration:
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this
device may not cause harmful interference, and (2) this device must accept any interference received, including
interference that may cause undesired operation. This device contains licence-exempt transmitter(s)/receiver(s) that
comply with Innovation, Science and Economic Development Canada’s licence-exempt RSS(s).
This device complies with the FCC and ISED Canada portable RF exposure requirements for an uncontrolled environment
and is safe for the intended operation as described in this manual.
Radiated output power (RF):
2.4 Ghz, 2.5mW
Rated Outputs / Voltages / Currents: Type 530
The electrical output of
PULSE
depends on the parameter settings of the active NMES programme. The programme
loaded at manufacture is described as “Essential Toning” and delivers the following output:
Parameter
500
!
1K
!
Peak Output Current (user adjustable)
0 - 80 mA
0 - 80mA
Peak Output Voltage
0 - 40V
0 - 80 V
Output Frequency range
70 Hz
70 Hz
Phase Duration
300 µS
300 µS
DC Component
none
none
The
PULSE
app allows you to select from a range of other NMES programs, each having different parameter selections.
The following table gives the limits of these parameters. All
PULSE
programs are within these limits.
Parameter
500
!
1K
!
Peak Output Current (user adjustable)
0 - 80 mA
0 - 80mA
Peak Output Voltage
0 - 40V
0 - 74.4V
Output Frequency range
4 -70Hz
4 -70Hz
Phase Duration
100-450 µS
100-450 µS
DC Component
none
none
The parameters of each programme can be viewed on the app. The maximum output power of the device into a 500
!
load is 180mW.
EU Only
• Wireless communication equipment such as wireless home network devices, mobile phones (other than your own),
cordless telephones, base stations and walkie-talkies of between 1W to 10W can affect this equipment and should be
kept a distance of least 2.3 to 7.3 metres away from the device.
19
English
Pulse IM Rev E.qxp_ML Book 27/06/2019 12:24 Page 20