Instructions – MEV
15
7.4 AUTOMATIC DEFROST, DISINFECTION AND POWER SAVING SETTINGS
Press
to open the Settings menu.
>
Chamber temp
-20 °C
Quickfreeze
Off
UV disinfection
Light
Auto
Select Program settings and press
to open the
menu for setting automatic defrost and power saving
times.
TIME TUE 13:05:05
Language
English
Display version
>
Program settings
5 µm
Enter PIN Code.
PIN Code is usually restricted to service technicians or
instrument administrators
Enter PIN
0000
Select defrosting programs 1 to 3 and set a time for the
automatic defrosting programme.
(e.g. each day at 12:00 h and at 23:00 h in this
example)
>
Defr. Start 1
12:00
Defr. Start 2
23:00
Defr. Start 3
__:__
Disinf. Req.
__:__
Select temperature increase start and set a starting
time for the automatic power saving mode.
(e.g. start at 21:00 h in this example)
Disinf. delay
10 h
>
Incr. start
21:00
Incr. end
04:00
Incr. days
____56_
Select temperature increase end and set a ending time
for the automatic power saving mode.
(e.g. end at 04:00 h in this example)
Disinf. delay
10 h
Incr. start
21:00
>
Incr. end
04:00
Incr. days
____56_