System Failure N
N = system failure
number which can be
between 1-255.
System Failure
Latching HIGH priority alarm.
Cannot be disabled.
Priority: HIGH
Type: TECHNICAL
Single or multiple sub-
system failures.
Remove patient to an
alternative form of
ventilation. Remove the unit
from service.
Gas Flow Error
Start up self test has detected
abnormal pneumatic system
response, or lacking gas supply.
Latching HIGH priority alarm.
Priority: HIGH
Type: TECHNICAL
Self test is repeated and
clears itself upon successful
result.
If ventilating - remove
patient to an alternative
form of ventilation. Remove
the unit from service.
High Pressure
Proximal pressure exceeds 15.5
mbar.
Non-latching HIGH priority alarm.
Priority: HIGH
Type: TECHNICAL
Cuts fresh gas flow and
evacuates pressure from
the circuit. 1.5 l/min bypass
flow is delivered, to flush
CO
2
build up. Device will
restart after 5 seconds.
Check Patient, check
patient circuit. Adjust
ventilation parameters.
If alarm repeats after the
above checks remove
patient to an alternative
form of ventilation. Remove
the unit from service.
Apnoea
Breath not detected within the user
set apnoea period.
Non-latching HIGH priority alarm.
Is reset by a patient breath.
Priority: HIGH
Type: PATIENT
Triggers a stimulation
breath of 3 mbar above set
CPAP for 0.3 seconds (only
pressure mode).
User has option for
disabling the apnoea
monitoring function (value
shows ‘--’).
Check Patient, check
patient circuit. Adjust
ventilation parameters.
Alarm message
Description
Action
Summary of Contents for SLE1000
Page 1: ...SLE1000 User manual Version 2 6 4 software ...
Page 4: ...Page 4 of 128 This page is intentionally blank ...
Page 8: ...Page 8 of 128 This page is intentionally blank ...
Page 9: ...Page 9 of 128 Introduction ...
Page 12: ...Page 12 of 128 This page is intentionally blank ...
Page 13: ...Page 13 of 128 SLE1000 Description ...
Page 17: ...Page 17 of 128 Operational and Clinical Warnings ...
Page 23: ...Page 23 of 128 User interface ...
Page 36: ...Page 36 of 128 Push the knob to select ...
Page 54: ...Page 54 of 128 This page is intentionally blank ...
Page 55: ...Page 55 of 128 CPAP Flow mode ...
Page 68: ...Page 68 of 128 This page is intentionally blank ...
Page 69: ...Page 69 of 128 SLE1000 Set up ...
Page 79: ...Page 79 of 128 Functional Testing ...
Page 85: ...Page 85 of 128 Operational Considerations ...
Page 87: ...Page 87 of 128 Operation ...
Page 89: ...Page 89 of 128 Troubleshooting Chart ...
Page 92: ...Page 92 of 128 This page is intentionally blank ...
Page 93: ...Page 93 of 128 Technical Information ...
Page 120: ...Page 120 of 128 This page is intentionally blank ...
Page 121: ...Page 121 of 128 Consumables Accessories ...
Page 123: ...Page 123 of 128 Glossary ...
Page 128: ......