No Air Supply
Air supply is too low (<2.0 bar) to
assure normal operation.
Non-latching HIGH priority alarm.
Suspends high oxygen alarm.
Priority: HIGH
Type: TECHNICAL
Device will deliver 100%
oxygen as set flow.
Connect air supply. If
ventilating - remove patient
to an alternative form of
ventilation. Remove the unit
from service.
No Oxygen Supply
Oxygen supply is too low (<2.0 bar)
to assure normal operation. Only
active if FiO
2
is set >21%.
Non-latching HIGH priority alarm.
Suspends low oxygen alarm.
Priority: HIGH
Type: TECHNICAL
Device continues to deliver
air only gas to patient.
User can prevent
recurrence by setting FiO
2
to 21%.
Connect oxygen supply.
If no oxygen present, set
the unit to run on air only by
canceling the alarm
message.
If ventilating - remove
patient to an alternative
form of ventilation. Remove
the unit from service.
High BPM rate
The measured breath rate exceeds
90 BPM for at least 6 seconds
when the Apnoea is set in the
range of 5 and 90 seconds.
Disabled if Apnoea is OFF (--)
High rate could indicate false
breath detection due to pressure
signal noise (e.g. from water in
breathing circuit). False detection
risk masking an apnoea condition.
Non-latching HIGH priority alarm.
Priority: HIGH
Check Patient, check
patient circuit. Adjust
Apnoea parameters.
Alarm message
Description
Action
Summary of Contents for SLE1000
Page 1: ...SLE1000 User manual Version 2 6 4 software ...
Page 4: ...Page 4 of 128 This page is intentionally blank ...
Page 8: ...Page 8 of 128 This page is intentionally blank ...
Page 9: ...Page 9 of 128 Introduction ...
Page 12: ...Page 12 of 128 This page is intentionally blank ...
Page 13: ...Page 13 of 128 SLE1000 Description ...
Page 17: ...Page 17 of 128 Operational and Clinical Warnings ...
Page 23: ...Page 23 of 128 User interface ...
Page 36: ...Page 36 of 128 Push the knob to select ...
Page 54: ...Page 54 of 128 This page is intentionally blank ...
Page 55: ...Page 55 of 128 CPAP Flow mode ...
Page 68: ...Page 68 of 128 This page is intentionally blank ...
Page 69: ...Page 69 of 128 SLE1000 Set up ...
Page 79: ...Page 79 of 128 Functional Testing ...
Page 85: ...Page 85 of 128 Operational Considerations ...
Page 87: ...Page 87 of 128 Operation ...
Page 89: ...Page 89 of 128 Troubleshooting Chart ...
Page 92: ...Page 92 of 128 This page is intentionally blank ...
Page 93: ...Page 93 of 128 Technical Information ...
Page 120: ...Page 120 of 128 This page is intentionally blank ...
Page 121: ...Page 121 of 128 Consumables Accessories ...
Page 123: ...Page 123 of 128 Glossary ...
Page 128: ......