Sampling

Allow pump to equilibrate after moving it from one temperature extreme 
to another.

1.  Replace the representative sample media used for 

calibration with new, pre-weighed media (

see Media 

Preparation and Impactor Preparation

).

2.  Mount the bracket at the desired location and at breath-

ing zone height (6 feet or 2 meters) using wire ties or 
other fasteners. Mount the impactor on the mounting 
bracket by threading the clamp knob into the bo

tt

 om of 

the impactor.

3.  Insert the screw on the rain cover into the screw hole in 

top of the impactor inlet and rotate the cover until tight.

4.  Turn on the pump and record sample start time, ambient 

temperature, ambient pressure, and other pertinent data.

Sample start time and duration can be programmed 
into the Leland Legacy Sample Pump in advance and 
sampling may be started manually or automatically.

Record all necessary data before disconnecting the 
pump from the ba

tt

 ery and reconnecting to the new 

ba

tt

 ery.

5.  After desired sample time has elapsed, turn o

  the pump and record the 

sample stop time, total volume, ambient temperature, ambient pressure, and 
other pertinent data.

6.  Remove the rain cover, reinstate the calibration 

train and media, and verify the 

fl

 ow rate (

see Cali-

bration

).

7.  Reach inside the case and press quick-connect 

release while pulling the tubing from the inlet. 
Remove tubing from the impactor. 

Technical Tidbits 

• 

The supplied rain cover should be used for all outdoor sampling.

• 

Keep the Leland Legacy sample pump inside the Pelican case and the 
case closed during sampling to protect the sample pump from weather.

3

4

7

Quick-connect 

release inside 

case

Tubing with 

quick-connect 

fi

 tting

Summary of Contents for DPS

Page 1: ...Operating Instructions SKC Deployable Particulate Sampler DPS System SKC Inc 863 Valley View Road Eighty Four PA 15330 Form 38049 Rev 1702 ...

Page 2: ...p flow rate with calibrator Record flow rate Reset accumulated data if required 4 Disconnect calibrator and remove calibration adapter from impactor Remove calibration media and place new unexposed media in impactor 5 Mount bracket at desired location 6 Screw impactor onto mounting bracket 7 Screw rain cover onto impactor Sampling 1 Turn on pump and record pertinent data Leland Legacy pump may be ...

Page 3: ...pler 6 Sample Pump Operation 7 Charging the Battery 7 Reading the Charging Status LED 8 Battery Setup 8 Battery Replacement 9 Leland Legacy Quick Guide 10 Calibration and Sampling 11 Calibration 11 Sampling 12 Sample Removal Shipping and Analysis 13 Removing the Collection Filter and Impaction Disc 13 Shipping Samples 13 Analysis 13 Ordering Information 14 Li Ion Battery Shipment 16 Warranty 16 In...

Page 4: ...e cut point are collected on a 47 mm filter The IMPACT Sampler mounts easily and makes media changes fast and simple with its convenient removable filter cassette and opener U S Patent No 7 334 453 The SKC DPS System includes a Leland Legacy Sample Pump with connection case andcharger 100 240V IMPACTsamplinghead 2filtercassettes calibrationadapter raincoverforsamplinghead 25disposableimpactiondisc...

Page 5: ...0 to 45 C Storing 4 to 95 F 20 to 35 C Operating 32 to 104 F 0 to 40 C Operating Humidity 0 to 95 non condensing Altitude Do not use pump beyond 7500 ft RFI EMI Shielding CE marked Case Dimensions 18 5 x 14 1 x 6 9 in 47 x 36 x 18 cm Complete System Weight 13 lbs 5 9 kg Sampling Head Impactor Dimensions 2 6 dia X 1 8 H x 3 8 L in 7 x 5 x 10 cm Sampling Head Impactor Weight 0 5 lb 0 23 kg Note The ...

Page 6: ...asses through the openings in the top of the filter cassette and follows through to a 47 mm filter where the smaller particles collect The top of the filter cassette accommodates the impaction disc For optimum impac tor performance a 37 mm oiled porous plastic disc is recommended as disposable collection substrate supplied with system If chemical analysis of larger particles is desired a 37 mm fil...

Page 7: ...aust and filter cas sette should be cleaned after five runs or upon a noticeable buildup of material This will remove oil buildup from the top of the filter cassette and other residue built up from frequent sampling Disassemble the impactor and wash parts in water with a liquid detergent or soap Rinse and air dry all parts thoroughly in a clean environment Do not place any mechanical object in the...

Page 8: ... cassette horizon tally into the U of the opener until the two halves of the cas sette loosen Gently pull halves apart Ensure the stainless steel support screen is in place in the bottom of the filter cassette Using forceps place a preweighed 47 mm filter on the support screen Press the filter cassette top into the filter cassette bottom Reinsert the cassette into the impactor 1 7 Technical Tidbit...

Page 9: ...r cassette see Inserting a Collection Filter into the IMPACT Sampler 1 Place the impaction disc in the recessed area on the filter cassette top The rough side of the impaction disc should face up SKC supplied discs are stamped with UP on the appropriate side Place the filter cassette on the exhaust and screw the impactor inlet and exhaust together just until tight Further hand tighten by 1 4 turn ...

Page 10: ...ot open disassemble short circuit crush incinerate or expose the battery to fire or high temperatures Tampering with the battery pack voids any warranty Ensure proper orientation of charging cable before plugging it into the charging jack Improper orientation contact will short circuit the battery and voids any warranty Short circuiting the battery pack will render it immediately inoperative Failu...

Page 11: ...LED steadily for 5 seconds to read charge status LED Action Charge Status ON steady Charge in progress ON 2 sec OFF 25 sec ON 2 sec Repeats Approximately 80 charged OFF 2 sec ON 25 sec OFF 2 sec Repeats Charge completed Battery Setup 1 Insert the plug on the connecting cable from the pump into the jack on the battery adapter on top of the external battery assembly 2 Insert the external battery ass...

Page 12: ...cable from the jack on the battery adapter on top of the external battery assembly 3 Insert the plug on the connecting cable into the battery adapter jack on a new fully charged external battery assembly 4 Insert the external battery assembly into the foam compartment in the case Ensure there is no tension on the connecting cable ...

Page 13: ...e new setting and place pump in Hold For CalChek Calibration see operating instructions To change temperature scale from F to C or C to F Press then Press until temperature displays Press or to switch units press until End displays then press to save new setting To change atmospheric pressure scale mm mb In Press then Press until pressure displays then press or to switch units press until End disp...

Page 14: ...e flow rate until the calibrator displays 10 L min see Leland Legacy Quick Guide Record the flow rate Reset the accumulated data before sampling See pump and calibrator operating instructions 1 When calibration is completed disconnect the calibrator and tubing from the calibration adapter Remove the calibration adapter from the impactor Remove the representative media and replace with fresh media ...

Page 15: ...on can be programmed into the Leland Legacy Sample Pump in advance and sampling may be started manually or automatically Record all necessary data before disconnecting the pump from the battery and reconnecting to the new battery 5 After desired sample time has elapsed turn off the pump and record the sample stop time total volume ambient temperature ambient pressure and other pertinent data 6 Remo...

Page 16: ...ee Ordering Information Accessories for glass jars Gently lift the filter cassette from the exhaust Use forceps to remove the collection filter and place in an appropriate container for shipping to a laboratory Unscrew the impactor inlet from the exhaust 1 SAMPLE REMOVAL SHIPPING AND ANALYSIS Removing the Collection Filter and Impaction Disc Shipping Samples Package and transport samples and blank...

Page 17: ...essories TSI 4146 Calibrator Kit 0 01 to 20 L min includes calibrator soft sided case mounting lugs tubing 1 4 inch ID battery pack 6 AA batteries inlet filter dampening module NIST certificate and manual 770 4146 Filter Cassette 225 396 Forceps stainless steel 225 8371 Petri Dish Slide for filter transport pk 100 225 2 01 Glass Jars for Chemical Analysis pk 8 pk 36 225 8376 225 8377 DataTrac Soft...

Page 18: ...attery Assembly with battery adapter 223 247 Battery Adapter for replacement of adapter on external battery assembly Cat No 223 247 223 248 Battery Connection Case with cable and plug 223 249 DPS DVS Charger Adapter 223 245 DPS DVS Charging Unit P22300 About Li Ion Battery Shipment Rechargeable lithium ion batteries for use with SKC sample pumps have been tested in accordance with the UN Manual an...

Page 19: ...y SKC products are subject to the SKC Limited Warranty and Return Policy which provides SKC s sole liability and the buyer s exclusive remedy To view the com plete SKC Limited Warranty and Return Policy go to http www skcinc com warranty asp ...

Reviews: