AirLite Quick Start Guide
On/Off
1. Lift control cover on top of pump.
2. Press on/off button.
Setting Flow Rate (1000 to 3000 ml/min)
1. Set up sampling train.
2. Connect calibrator to inlet of sample medium.
3. Turn on pump and adjust flow adjustment screw on pump until
desired flow rate is indicated on calibrator. Take a minimum of
three readings. Calculate flow as the average of three readings.
LED Indicators
Slow flash
= normal run
Fast flash
= flow fault
Double flash
= low battery, approximate minimum one-hour run time
remaining
Flash every eight seconds
= pump shutdown due to exhausted batteries
Four flashes every six seconds
= pump shutdown due to flow fault
Note:
Indication of pump shutdown due to exhausted batteries will override all
other indicators.
AirLite uses three replaceable standard AA alkaline batteries.