AirLite Quick Start Guide

On/Off

 

1. Lift control cover on top of pump.

 

2. Press on/off button.

Setting Flow Rate (1000 to 3000 ml/min)

 

1. Set up sampling train.

 

2. Connect calibrator to inlet of sample medium.

 

3. Turn on pump and adjust flow adjustment screw on pump until  

 

  desired flow rate is indicated on calibrator. Take a minimum of  

 

  three readings. Calculate flow as the average of three readings.

LED Indicators

 

Slow flash

 = normal run

 

Fast flash

 = flow fault

 

Double flash

 = low battery, approximate minimum one-hour run time 

 

remaining

 

Flash every eight seconds

 = pump shutdown due to exhausted batteries

 

Four flashes every six seconds

 = pump shutdown due to flow fault

Note:

 

Indication of pump shutdown due to exhausted batteries will override all 

other indicators.

AirLite uses three replaceable standard AA alkaline batteries.

Summary of Contents for AirLite 110-100

Page 1: ...Operating Instructions AirLite Sample Pump Catalog No 110 100 SKC Inc 863 Valley View Road Eighty Four PA 15330 Form 37814 Rev 0802 ...

Page 2: ......

Page 3: ...al Accessories 13 Service Policy 14 Warranty 15 Notice This operating instruction may not address all safety concerns if any associated with this product and its use The user is responsible for determining and following the appropriate safety and health practices and regulatory limitations if any before using the product The information contained in this document should not be construed as legal a...

Page 4: ......

Page 5: ...ator Take a minimum of three readings Calculate flow as the average of three readings LED Indicators Slow flash normal run Fast flash flow fault Double flash low battery approximate minimum one hour run time remaining Flash every eight seconds pump shutdown due to exhausted batteries Four flashes every six seconds pump shutdown due to flow fault Note Indication of pump shutdown due to exhausted ba...

Page 6: ... ...

Page 7: ...3000 ml min flow range with full back pressure compensation Features Economical alkaline battery operation Built in pulsation dampener Built in replaceable particulate trap Easy access battery compartment Impact resistant RFI shielded case Front View User option auto restart from flow fault One year limited warranty Economical alkaline battery operation LED indicates pump running flow fault and lo...

Page 8: ...or disposable batteries Case Plastic RFI EMI shielded CE marked LED Functions Slow flash normal run Fast flash flow fault Double flash low battery approximately one hour run time remaining Flash every eight seconds pump shutdown due to exhausted batteries Four flashes every six seconds pump shut down due to flow fault Battery run time Dependent on batteries used See Table 1 Dimensions 4 6 x 3 x 1 ...

Page 9: ... mm 25 mm 25 mm Flow Rate 1 L min 2 L min 2 5 L min 1 L min 2 L min Duracel Standard 23 5 14 5 13 5 18 5 10 5 Rayovac Maximum 20 0 16 5 14 0 16 5 11 0 Wal Mart EverActive 24 0 16 0 10 5 18 0 10 5 Eveready Alkaline 20 0 14 0 13 0 17 0 Not tested Polyvinyl Chloride PVC Filter 0 8 µm pore size Filter Diameter 37 mm 37 mm 37 mm 25 mm 25 mm Flow Rate 1 L min 2 L min 2 5 L min 1 L min 2 L min Duracel St...

Page 10: ...f the case close the door and tighten the screw until secure Important Note the polarity of the batteries Note For maximum run time insert new batteries in pump before sam pling If using rechargeable AA 1 2 volt NiMH batteries expect approximately half the run time stated for disposable batteries Note To prevent corrosion of battery terminals remove batteries when AirLite will not be used for an e...

Page 11: ...e 1 6 clockwise to increase the flow or counterclockwise to decrease the flow until the desired flow rate is indicated on the flowmeter When the desired flow is obtained take three readings to confirm that the flow rate is consistent Calculate and record the flow rate as the average of three readings Disconnect the flowmeter and replace the representative sampling medium with a new unexposed mediu...

Page 12: ...C marked to sample to the Adjustable Low Flow Tube Holder Figure 2 Label all tubes and ports Insert opened representative tubes into the rubber sleeves of each port on the Adjustable Low Flow Tube Holder If any port remains unused place an unopened tube in it it is important to seal unused ports Use tubing to connect the exposed end of one tube to a primary standard flowmeter Loosen the screw on t...

Page 13: ...recautions when using these tubes Failure to use the necessary traps will damage the pump and void the warranty Caution Do not operate AirLite in hazardous or explosive locations AirLite is designed for applications that do not require intrinsic safety Note The CPC has two small inlet ports on the bottom of the unit These ports should be periodically inspected for blockage which can occur when sam...

Page 14: ...is Caution Do not operate AirLite in hazardous or explosive locations AirLite is designed for applications that do not require intrinsic safety Caution Protect sample pump from weather when in use outdoors Reading the LED The LED on top of the pump case Figure 1 7 indicates pump and battery status Slow flash Normal run Fast flash Flow fault Double flash Low battery approximately one hour run time ...

Page 15: ...tries Disabling the Fault Shutdown and Auto restart Features If the fault link is fitted when the pump is initially switched on the LED will display a quick double flash before running If the link is not fitted the pump will immediately run without the LED flashing To disable the fault shutdown and auto restart features remove the two screws from the belt clip and set the clip and screws aside Rem...

Page 16: ... Inlet hose connect P20106 3 Pulsation dampener P2010802 4 Stack plate N A 5 Valve plates top and bottom P213201 6 Diaphragm Yoke assembly P22417H 7 Pump body P22417G 8 Motor Eccentric P51890 9 Pump base P20182 Battery case cover P20184 not shown 1 5 1 4 9 2 6 8 7 3 5 ...

Page 17: ...ded in Low Flow Adapter Kit above 224 26 01 Dual Adjustable Low Flow Tube Holder 224 26 02 Tri Adjustable Low Flow Tube Holder 224 26 03 Quad Adjustable Low Flow Tube Holder 224 26 04 Protective Cover for Sorbent Tubes Type A included in Low Flow Adapter Kit above 224 29A Protective Cover for Sorbent Tubes Type B 224 29B Protective Cover for Sorbent Tubes Type C 224 29C Protective Cover for Sorben...

Page 18: ...ress SKC Inc National Service Center 863 Valley View Road Eighty Four PA 15330 Package product carefully to prevent damage during transit Include a contact name phone number shipping address RMA number and a brief description of the problem For nonwarranty repairs a purchase order number and billing address are also required The Service Department will contact nonwarranty customers with an estimat...

Page 19: ...rty damages for loss of use loss of time loss of profits loss of income or other incidental damages Nor shall buyer be entitled to recover from SKC Inc any consequential damages resulting from defect of the instrument including but not limited to any recovery under section 402A of the Restatement Second of Torts 3 This warranty extends only to the original purchaser of the warranted instrument dur...

Reviews: