Form 37182 Rev 201019

Page 1 of 12

AirLite Sample Pump with Timer

Cat. No. 110-100-T

Operating Instructions

Figure 1.

 AirLite Sample Pump with Timer

863 Valley View Road, Eighty Four, PA 15330  USA • Tel: 724-941-9701 • www.skcinc.com

Accessory 

mounting 

screws

Protective 

control cover

 

(raised)

Belt clip

 (back of pump)

Battery compartment

Pump inlet 

(inlet filter 

underneath)

Bottom view

Top view

Battery compartment door

Flow 
adjustment 
screw

On/off 
button

Battery 
compartment 
screw

Hinge

Introduction

Description

The AirLite Sample Pump with Timer 

(Figure 1)

 provides 5 to 3000 ml/min flows and is suitable for abatement, indoor air sampling, 

and emergency response in non-hazardous locations.

Checking Pump/Kit Contents

Use the table below to verify that you received all items associated with the Cat. No. ordered. If you are missing items, contact  
SKC at 800-752-8472 (U.S. only) or 724-941-9701.

If You Ordered Cat. No. Your Package Should Contain
110-100-T

AirLite Sample Pump with Timer only, with three AA alkaline batteries and screwdriver set

110-100-TK5

High Flow Sample Pump Kit includes 5 pumps as described above and filter cassette holders, in a Pelican case

110-100-TK5D

High/Low Flow Sample Pump Kit includes 5 pumps as described above, filter cassette holders, All-in-One adjustable 
tube holders, and Type A protective tube covers, in a Pelican case

Required Equipment

 

1/4-inch ID tubing

 

Low flow accessories

 if sampling at 5 to 500 ml/min.

 See Accessories.

Front view

LCD

Summary of Contents for AirLite 110-100-T

Page 1: ...and emergency response in non hazardous locations Checking Pump Kit Contents Use the table below to verify that you received all items associated with the Cat No ordered If you are missing items contact SKC at 800 752 8472 U S only or 724 941 9701 If You Ordered Cat No Your Package Should Contain 110 100 T AirLite Sample Pump with Timer only with three AA alkaline batteries and screwdriver set 110...

Page 2: ...ent door under the lip of the case close the door and tighten the screw until secure Notes and Cautions Do not operate AirLite with Timer in hazardous or explosive locations AirLite with Timer is designed for applications that do not require intrinsic safety Failure to follow warnings or cautions voids any warranty For maximum run time insert new batteries in pump before each sampling period If us...

Page 3: ...calibrator outlet suction port to the representative sample medium inlet Using 1 4 inch ID tubing connect the representative sampling medium outlet to the pump inlet 4 Using a small screwdriver set the pump flow rate by turning the flow adjustment screw Figure 1 clockwise to increase the flow or counterclockwise to decrease the flow until the calibrator indicates the method specified flow rate Tak...

Page 4: ...is unable to compensate due to excessive back pressure or a low battery condition exists it will shut down and timing functions will pause The LCD will display either a battery shaped icon or a flow fault icon depending on the cause of the shutdown Upon flow fault the pump will attempt to restart up to five times To restart from flow fault correct the blockage and press the on off button twice The...

Page 5: ...e tubing from the pump inlet Flow Rate Through Sorbent Tube 1 Attach the Tygon tubing of theAll in One to the pump inlet See calibration train at right Note If the pump flow faults shortly after the holder is attached check that the needle valve is open by using a small flat head screwdriver to turn the flow adjust screw on the holder slightly counterclockwise If flow fault continues check that th...

Page 6: ...information Possible Displays During Sampling Flow or Battery Fault Shutdown If the pump is unable to compensate due to excessive back pressure or a low battery condition exists it will shut down and timing functions will pause The LCD will display either a battery shaped icon or a flow fault icon depending on the cause of the shutdown Upon flow fault the pump will attempt to restart up to five ti...

Page 7: ...mber by 0 15 and total the two numbers Set the pump flow rate for the resulting new sum Example Sampling with three sorbent tubes each with a flow rate of 500 ml min The sum of the tube flow rates is calculated as 3 x 500 1500 Determining a 15 higher flow rate is calculated as 1500 x 0 15 225 Calculating the final pump flow setting would be 1500 225 1725 ml min Back pressure across some sample tub...

Page 8: ... for blockage that can occur when sampling in dusty environments Such blockage may cause back pressure to increase Clean ports with a small pick and blow away any particles with a puff of air 1 Replace the representative sorbent tubes used for calibration with new unexposed sorbent tubes for sample collection 2 Place a protective tube cover over each tube and thread into place on holder until secu...

Page 9: ...w a Turn on the pump and reinstate the calibration train and sample media b Take three readings and record the average value as the post sample flow rate Do not adjust the pump flow rate during this step c Compare the pre and post sample flow rates Note in sampling documentation if the values differ by more than 5 Maintenance See Insert Replace the Batteries Pump Service Pumps under warranty shoul...

Page 10: ...ow Adapter Kit above 224 29A Type B 8 mm OD x 110 mm L 224 29B Type C 10 mm OD x 150 mm L 224 29C Type D 10 mm OD x 220 mm L 224 29D Miscellaneous Screwdriver Set included with pump 224 11 Protective Nylon Pouch with belt loop black 224 902 Waist Strap 224 12 Replacement Parts Cat No 1 Stack screws P51891 2 Inlet hose connector P20106 3 Pulsation dampener P2010802 4 Stack plate N A 5 Valve plates ...

Page 11: ...ble to pump compensation range to determine appropriate applications Temperature Ranges Operating 32 to 104 F 0 to 40 C Storage 4 to 113 F 20 to 45 C Compensated Temperature Range 41 to 122 F 5 to 50 C Operating Humidity 0 to 95 non condensing Power Three standard AA alkaline batteries Rechargeable AA size 1 2 volt NiMH batteries may be used but will provide approximately half the run time stated ...

Page 12: ...e size Filter Diameter 37 mm 37 mm 37 mm 25 mm 25 mm Flow Rate 1 L min 2 L min 2 5 L min 1 L min 2 L min Duracell Standard 23 5 14 5 13 5 18 5 10 5 Rayovac Maximum 20 0 16 5 14 0 16 5 11 0 Wal Mart EverActive 24 0 16 0 10 5 18 0 10 5 Eveready Alkaline 20 0 14 0 13 0 17 0 Not tested Polyvinyl Chloride PVC Filter 5 0 µm pore size Filter Diameter 37 mm 37 mm 37 mm 25 mm 25 mm Flow Rate 1 L min 2 L mi...

Reviews: