Troubleshooting
Should the Flite4 pump detect that the battery is at a low charge state,
a flat battery message will be displayed on the LCD screen. Charge the
battery fully using the correct SKC charger before use.
Flat Battery
CONNECT CHARGER
The Flite4 pump control board monitors the pump motor speed whilst
running. If the motor speed signal is not detected, for example due to
a disonnected wiring connection to the motor or a failure of the motor,
the pump will display a motor fault warning message and retain the
elapsed runtime in memory. Press the and keys simultaneously
to clear the message and return to the ‘SELECT’ main menu screen.
* Motor Fault *
001:17:48
+
Æ
SELECT Sleep
Run Setup Review
Possible Fault
Corrective Action
Sample media fault
1. Check media for heavy loading or damage.
2. Rectify problem.
3. Try restarting the pump.
Sample train fault
1. Check tubing for blockage or crimping.
2. Rectify problem.
3. Try restarting the pump.
Software error (frozen keypad)
1. Disconnect battery for one minute, reconnect.
2. Try restarting programming.
For further assistance contact your supplier or SKC customer care - +44 (0) 1258 480188.