24
Battery Backup
In case of power failure the emergency battery backup will operate the bed. Use only
in emergency situations, the battery backup is not designed for continuous use. The battery
will recharge itself while the bed is plugged into a hospital grade or grounded outlet.
New batteries, or batteries stored for a long time, may take more time to charge. Keep the
battery near room temperature when charging.
Features and Options:
Usage
:
The specifications and characteristics of the battery are from the manufacturer’s guidelines sheet.
If the battery needs to be changed, follow the manufacturer’s specifications listed in this section
as well as the information listed in the Specifications section (page 10).
•
Lithium Ion technology: Nickel Manganese Cobalt Oxide (NMC)
•
Output voltage: 25.9 V DC
•
Max. discharge current: 10 A
•
Charging Current: 300 mA/controlled by BA21
•
Battery capacity: 2.25 Ah/58.28 Wh
•
Weight: 0.7 kg
•
Housing colour: Light grey (RAL 7035)
•
Protection class: IPX6 Washable DURATM
•
Packaging: Every battery is packed individually and is fitted with
lithium caution (transportation requirement)
•
Classification: Internally powered
•
Compatibility: Battery back-up for CO
61
•
Duty cycle: 5 %, 1 minute continuous use followed by
19 minutes not in use.
•
Charging: With integrated charger in battery
•
Charging time: Approx. 10 hours
•
Recharging during storage: First recharge of the battery must be
no
later than 12 months after production date stated on the label.
Hereafter the battery must be recharged at least every 12 months.
•
Operating temperature: + 5 °C to + 30 °C
•
Storage temperature: - 10 °C to + 40 °C
(+ 10 °C to + 25 °C recommended) The batteries must be stored
in an
applicable storage room without direct sunlight
•
Relative humidity: 20 % RH to 80 % RH, non-condensing
•
Atmospheric pressure: 700 to 1060 hPa
•
Approvals: IEC60601-1, ANSI/AAMI ES60601-1
CAN/CSA-22.2 No 60601-1, IEC62133, UL2054,
UN38.3 (needed for transport of lithium batteries)
BA-21
Lithium Ion