4 Installing and connecting the unit
Operating Instructions SIROTorque L
+
60 52 208 D 3489
D 3489.201.01.03.02
11
båÖäáëÜ
Supplementary notes for installing
SIROTorque L
+
The connecting cable (7) between the transformer (12)
and the SIROTorque L
+
(1) must be run in such a way
that it can
not
be damaged by moving appliances such
as chairs with wheels, carts, etc.:
•
Place the transformer (12) directly next to the supply
unit (C).
•
Run the connecting cable (7) so that it does not
touch the floor anywhere. To do this, use the
adhesive cable clamps (D) supplied.
•
Run the connecting cable (7) in such a way that it
cannot be damaged at the supply unit.
•
Install the power cable (13) of the transformer as
required by the relevant national regulations.
Satellite version
Upon your request, the supplied SIROTorque L
+
table-
top unit may be converted to a "satellite version" by your
service engineer.
Conversion kit, Order No: 60 57 223
In this version, the control panel is separated from the
SIROTorque L
+
tabletop unit and is connected to the
latter by a permanently wired cable having a length of
approx. 3 m.
The control panel (A) can be fastened at a place that is
easily accessible for the user, e.g. in the control panel
front area, either with screws (from the top) or with the
Velcro strap included in the scope of supply (from the
front).
Fastening of the control panel is done at your own risk,
make sure not to damage any cables etc. with the
screws.
The SIROTorque L
+
supply unit (B) can be installed
e.g. below the dentist element.
Establish all connections as described under
"Connection".
ATTENTION
Repair and maintenance work on the SIROTorque L
+
must be performed only by properly trained service engi-
neers. Before opening the housing, pull the power plug or
disconnect all poles from the power supply so as to deen-
ergize the unit.
W
ait for at least 10 seconds.
The compressed air delivered by the highspeed hand-
piece or air motor hose must be dry, oil-free and dirt-free.
It must not be contaminated!
If necessary, install a filter or water separator.
The spray water must be dirt-free according to the appli-
cable regulations. It must not be contaminated.
5
Siro
na
Dental
Syste
ms
Gm
bH
Fab
rikstraß
e 3
1
D-6
46
25
Ben
sheim
Germ
any
Sirona
Dental
Systems
GmbH
Fabrikstraße
31
D-64625
Bensheim
Germany
TIY
E24
-02
M (2
30V
)
TIYE24-02M
(230V)
TIY
M2
4-0
2M
(12
0V)
TIYM24-02M
(120V)
0123
120 V / 60 Hz
A
B
7
C
12
1
D
13
Summary of Contents for SUROTorque L+
Page 51: ......