4 Operation
Operating Instructions Simulation Unit SIM MODULAR
4.5 Dentist element
båÖäáëÜ
On the C+ electronic foot switch
✔
An instrument is removed from the holder.
➢ Press the left button of the C+ electronic foot switch.
When the status display
Spray
lights up on the EasyPad, spray is
switched on when the instrument is activated.
Activation of the spray is saved on the favorite keys, see section "Setting
the speed on the EasyPad" [ → 73], "Setting the intensity on the EasyPad"
[ → 90] and "Integrated torque control on the EasyPad" [ → 79].
4.5.6.1.2
Activating the chip blower
With the chip blower, an air jet comes out of the burr instrument nozzle.
✔
A highspeed handpiece or motor is removed from the holder.
➢ Press the right button of the pneumatic or C+ electronic foot switch.
The chip blower remains switched on as long as the button is
pressed.