62 14 972 D 3507
46
D 3507
.
031.01.07
.
02 11.2010
Electrical connection
Sirona Dental Systems GmbH
Connecting the release and the door contact
Installation Instructions HELIODENT
PLUS
6.4.9
Installation option 6.1
empty
Configuration
Installation option 6.1
Release outside of the X-ray room with remote control, door contact
safety circuit
● Door contact
– Door contact S7 wired to the remote control housing
● Release
– Manual release S3
or
– Remote control release key S9
As in installation version 6:
empty
Exception
Door contact wired via remote control.
empty
Prerequisites
● The remote control has been installed (see section "Installing the
remote control/remote timer [ ➙ 23]").
● Only with use of the manual release S3 (coiled cable):
– The holder for manual release S3 was attached to the cover of
the remote control. See section "Installation on the protective
cover of the remote control [ ➙ 29]".
6.1
L9
S7
S3
S9
NOTICE
Installation requirement
Use of the remote control is permissible only if the yellow X-Ray LED is
visible to the operating personnel during radiation release.