1 Preparations
Sirona Dental Systems GmbH
Installation Requirements C5
+
, C5
+
Turn
60 27 978 D3415
6
D3415.021.03.08.02
Identification of the danger levels
To prevent personal injury and material damage, please
observe the warning and safety information provided in the
present Operating Instructions. Such information is high-
lighted as follows:
DANGER
An imminent danger that could result in serious bodily injury
or death.
WARNING
A possibly dangerous situation that could result in serious
bodily injury or death.
CAUTION
A possibly dangerous situation that could result in slight
bodily injury.
NOTICE
A possibly harmful situation which could lead to damage of
the product or an object in its environment.
IMPORTANT
Application instructions and other important information.
Tip:
Information on making work easier.
General safety information
WARNING
The installation must be carried out in accordance with the
requirements stated in our Installationsvoraussetzungen.
WARNING
Installation may be carried out only by personnel specifical-
ly authorized by Sirona.
WARNING
For reasons of product safety, this product may be
operated only with original Sirona accessories or
third-party accessories expressly approved by
Sirona. The user is responsible for any damage re-
sulting from the use of non-approved accessories.
If any equipment not approved by Sirona is con-
nected, it must comply with the applicable stan-
dards, e.g.:
• IEC 60950-1 for information technology equip-
ment and
• IEC 60601-1 for medical electrical equipment.
In case of doubt, contact the manufacturer of the
system components.
WARNING
Any person who assembles or modifies a medical
electrical system complying with the standard
IEC 60 601-1 Chapter 16 by combining it with other
equipment (e.g. by connecting a PC) is responsible
for ensuring that the requirements of this regula-
tion are met to their full extent for the safety of the
patients, the operators and the environment.
NOTICE
The floor must be flat and level (DIN 18 202). A mounting
plate must be used for uneven floors (see Section 1.6,
"Mounting plates for equipment with fixed cuspidor (not for
C5
+
Turn)" on page 13).
CAUTION
The floor must have a minimum loading capacity of 0.5N/
cm².
CAUTION
Interference of electromedical devices caused by radio tele-
phones:
To ensure the operational readiness of electromedical de-
vices, the use of mobile radio telephones in the practice or
hospital area is prohibited.
CAUTION
Electromagnetic compatibility:
The unit should not be op-
erated in the immediate vicinity of other devices. If this
proves to be unavoidable, the unit should be monitored to
ensure that it is operating properly.
1.1 Safety