13 Water unit
Sirona Dental Systems GmbH
13.4 Disinfection system (optional)
Operating Instructions C4
+
59 06 313 D 3382
70
D 3382.201.01.13.02
13.4
Disinfection system (optional)
When the message
DESINF
flashes on the dentist ele-
ment display during treatment, you should refill 1 liter of
DENTOSEPT P
at the latest after the end of the treat-
ment day.
When the message flashes, it is
not
possible to perform
sanitation.
Remove the cover behind the cuspidor and fill in 1 liter
of
DENTOSEPT P
.
The
filler trough
should always be kept
clean
. Always
make sure that
no tap water
enters the filler trough!
For reorders:
DENTOSEPT P:
1 carton = 6 x 1 liter, REF 33 18 156
For Canada and Japan only
:
1 carton = 6 x 1 liter, REF 59 69 535
Microbiological water test
(after an operating pause of
more than
1 week
)
Sanitation of the unit
Perform sanitation every
4 weeks
. This is the only way
to effectively combat the formation of biofilm in the water
paths (see instructions
“Care and Cleaning by the
Practice Team”
).
DESINF
A
RPM