SIQENS Ecoport 800
Service and maintenance
Page 41 of 48
6.1.
Trouble shooting
In most cases, errors can be eliminated by simply checking the installation setup. Before the first
start, check the following points:
Battery cable
Description: The Ecoport does not perform a
function check (start the fans) after connecting
the battery and the signal light on the control
panel does not light up.
Solution: Make sure that the battery cable is
firmly seated and fully snapped into place.
Tank connection
Description: Air is drawn in via a connected tank.
In the control panel, this could be signaled -
among others - by error #48
(E_MAX_SYSTEM_RESTARTS).
Solution: Make sure that the dispensing head is
firmly connected to the attached cartridge or
external tank.
Solution: Make sure that the O-rings on the
dispensing head are correctly positioned
Air supply and exhaust
Description: The system overheats during
operation. In the control panel, this may be
indicated - among others - by error #21
(E_ME_TEMPERATURE_ CRITICAL).
Solution: Check the supply and exhaust air pipes
for blockages and ensure that the exhaust air is
correctly released to the outside via the exhaust
air pipe.