Sintrol DUMO User Manual Download Page 3

Page 3 

1. INTRODUCTION 

This manual describes how to install and use SINTROL digital triboelectric 
ambient dust monitor Dumo. 

Sintrol shall not be held liable for any loss or damage whatsoever arising 
from omissions or errors in this manual or any misuse of the product.  

1.1. Safety 

Dumo ambient dust monitor is virtually maintenance free. Under no 
circumstances should the user attemp to replace any components or the PC 
board. If, for some reason the monitor fails to operate, contact your local 
distributor or the manufacturer. 

It is essential that the unit is correctly grounded!  

(see section 3.)

1.2. Product overview 

Dumo is a dust monitor based on particle charge transfer. It utilises the 
experience and test data of the same technology Sintrol has developed and 
perfected over almost two decades. The device has been developed for 
continuous dust monitoring in industrial environment and it is equipped with three 
colour LED indicating different dust concentration levels. Dumo is designed for 
easy installation and operation. 

1.3. Operating principle 

  

Dumo measurement principle is based on the charge transfer of solid particles. 
When solid particles enter into the flow channel, they interact with a sensor rod 
and affect a small electrical charge. This charge provides a signal which is 
proportional to the ambient dust level. Schematic diagram of the measurement 
principle is presented in Figure 1. 

EASTERN ENERGY SERVICES PTE LTD.

60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG 415979 Singapore
Telephone: +65-6748 6911 Fax: +65-6748 6912
Email: [email protected] Website: 

www.eesgroup.com

Summary of Contents for DUMO

Page 1: ...User Manual s EASTERN ENERGY SERVICES PTE LTD 60 Kaki Bukit Place 02 19 Eunos Tech Park Singapore SG 415979 Singapore Telephone 65 6748 6911 Fax 65 6748 6912 Email info eesgroup com Website www eesgro...

Page 2: ...ATION 2 1 Selecting the installation location 3 WIRING 4 TECHNICAL SPECIFICATIONS 6 MAINTENANCE EASTERN ENERGY SERVICES PTE LTD 60 Kaki Bukit Place 02 19 Eunos Tech Park Singapore SG 415979 Singapore...

Page 3: ...ce and test data of the same technology Sintrol has developed and perfected over almost two decades The device has been developed for continuous dust monitoring in industrial environment and it is equ...

Page 4: ...of dust particles during the process If the user is interested in dust concentration level in general then Dumo should be located in a more peaceful place some distance from the process There are sti...

Page 5: ...4 20 mA red 8 Relay 2 output blue 7 Relay 1 output For Dumo It is extremely important that it is grounded properly The recommended grounding is presented below where Dumo cable shield is connected to...

Page 6: ...signals 2 solid state relays Isolated 4 20 mA Serial communication Ambient temperature 20 60 C Enclosure casing Aluminium Stainless steel Power supply 24 VDC Cable connection 5 meter cable Weight Abo...

Page 7: ...ffice air Alert level 1 000 000 Alarm level 4 000 000 In normal indoor environment Dumo signal level should stay under 5000 at the most of time Ambient monitoring and ambient dust concentration level...

Page 8: ...d way of monitoring When using this option it is possible to change alarm levels and averaging time for example The third possibility is to monitor data on line using the mA output supplied with a rem...

Page 9: ...urement channel after some period of time The cleaning period varies depending on monitoring conditions and dust concentrations The user should check the channel frequently EASTERN ENERGY SERVICES PTE...