Information for Healthcare Professional Use
·
Only trained healthcare professional be allowed to use Safe-Accu 2 blood glucose
monitoring system on multiple patients.
·
Only trained healthcare professional be allowed to collect venous blood samples.
·
Any patient who has infectious diseases must use his/her own meter.
·
If the same meter be used on multiple patients, there is risk of infection between both
patients and healthcare professional.
·
Water or other disinfection solution which remains on skin may dilute blood and cause
inaccurate test results.
·
Please dispose used lancet, test strips and expired control solution as medical waste.
Perform test:
1. Wear disposable gloves.
2. Healthcare professional use an alcohol swab to clean lance site. Then be sure dry stick
site before testing.
3. Use qualified lancing device and choose suitable penetrate depth for patients.
4. Using lancing device to lance.
5. Hold the tip of the test strip to the blood sample until the test strip reaction zone is full
and the meter begin countdown.
6. Wash hands thoroughly with soap and water after handling the meter, lancing device or
test strips.
7. Cleaning and disinfecting meter and lancing device before use on next patient.
Comparing Meter with Laboratory Results
When comparing results between Safe-Accu 2 blood glucose monitoring system and a
laboratory system, Safe-Accu 2 blood glucose monitoring system blood tests should be
performed within 30 minutes of a laboratory test.
Cleaning and Disinfecting
·
Clean and disinfect meter immediately after getting any blood on the meter or if meter is
dirty.
·
If the meter is being used on second person, the meter and lancing device should be
clean and disinfected first. Do not clean the meter during the test.
To clean the meter
1. Wash hands thoroughly with soap and water.
2. Make sure meter is off and a test strip is not inserted. Use 75% alcohol to rub the entire
outside of the meter. Make sure no liquids enter the test strip port or other opening in the
meter.
3. Let meter air dry thoroughly before using to test.
4. Wash hands thoroughly again after handing meter.
5. Verify the meter is working properly by performing control solution test. If your meter
can’t work properly, please contact customer service at +86-731-89935581/
+86-731-89935582 or contact local dealers.