SL-889 HDMI 2.1 Cat-3 Cable EMI Tester
User Guide
Simplay-UG-SL889-1.1
© 2019-2020 Simplay Labs, LLC.,
11
All rights reserved. CONFIDENTIAL
Setting up Connections
–
Pre-Test Operation
Connecting the DUT cable and optional HDMI monitor cable to the SL-889:
•
When performing EMI RF scans, the DUT cable should be connected from HDMI OUT1 to HDMI IN1 as shown in
•
When pre-testing, an optional 4K TMDS monitor can be connected to visualize the ITU BT1729 test pattern as
shown in
. Note: As the test is intended to measure only the radiated emissions of the DUT cable, the
cable from HDMI OUT2 should be disconnected from the SL-889, and the 4K Monitor should be unplugged from AC
power prior to making any EMI measurement scans.
Figure 2.1. Pre-Test: Connecting the DUT cable with optional monitor output