User manual for flow meter with pulse input SPI-94
controlled due to total flow counter result. It can be set to one of two values
:
“lit.”
- Litres
“m 3”
- cubical meters
Parameters ”
t PrEc
” and ”
t unit
” are available only if total flow counter is used to
control active current output (
“vALUE”
=
”
tot
”). When current flow rate is used to
control it, the unit and precision of
”OUt LO”
and
”OUt HI”
parameters are
defined by parameters
“F unit”
,
“Ftunit”
and
“F PrEc”
of menu
”Flouu”
.
”
Lo r
”, ”
Hi r
”
- this parameters define the output current range (maximum range 0 to 24 mA).
If calculated output value I
out
exceeds defined range then current output
generates the current equal to upper or lower border of the defined range. The
parameters define the percentage extension of nominal current range 4-20 mA
(with 0,1% resolution).
Parameter “
Lo r
” defines lower border of the range due to formula:
I
min
= 4 mA - 4 mA × “
Lo r
” %.
This parameter can be set from 0 to 99.9%.
Parameter
“Hi r”
defines lower border of the range due to formula:
I
max
= 20 mA + 20 mA × “
Hi r
” %.
This parameter can be set from 0 to 19.9%.
In example on page 38 of the
DISPLAY VALUES CALCULATION
paragraph
the procedure of the current outputs determining is presented in details.
“
AL
”
- this parameter determines the behaviour of current output if any critical situation
occurs. The parameter can be set to one of the values:
“noCHAn”
- current will not change,
“22.1”
- current will be set to 22.1 mA,
“3.4”
- current will be set to 3.4 mA,
“0.0”
- current will be set to 0 mA.
When the critical situation goes, the current will be set to value calculated due to
formulas given above.
If parameter
“OUtmod”
is set to
“on”
,
“oFF”
,
“0-20”
or
“4-20”
, the “critical situation”
means that
permissible measurement range
is exceeded.
If parameter
“OUtmod”
is set to
“modbuS”
, the “critical situation” means
communication delay (when no data is received) longer than
“mbtimE”
parameter (see
description:
“rS-485” menu
).
7.3.9.
”briGHt”
parameter
This parameter allows user to set bright of the LED display, bright can be set to
conventional values from 1 to 8.
32