© SIKA • Ba_TP17-TPM-Multi_en.doc • 01/2014
14.4 Characteristics series TP M … S
Series
Characteristics
TP M 165 S
TP M 225 S
TP M 255 S
Calibrator
Display range
-50…165 °C
0…225 °C
0…255 °C
Setting
range:
- Dry block
-35…165 °C
T
R
…225 °C
T
R
…255 °C
- Infrared black body
-35…165 °C
T
R
…225 °C
T
R
…255 °C
- Surface temperature
-25…150 °C
T
R
…200 °C
T
R
…200 °C
- Micro calibration bath(water)
0…100 °C
T
R
…100 °C
T
R
…100 °C
- Micro calibration bath (silicone
oil)
-35…165 °C
T
R
…225 °C
T
R
…255 °C
Tolerance:
- Dry block
0,3 °C
0,4 °C
0,4 °C
- Infrared black body
0,5 °C
0,5 °C
0,5 °C
- Surface temperature
1,0 °C
1,0 °C
1,0 °C
- Micro calibration bath
0,1 °C
0,2 °
0,2 °
Emission degree
- Infrared black body
0,9994
Control stability:
- Dry block
0,05 °C
- Infrared black body
0,05 °C
- Surface temperature
0,2 °C
- Micro calibration bath
0,05 °C
Electrical Characteristics
Power
supply:
- 90…240 VAC +/- 10 %
50/60 Hz
-/-
-/-
- 230 VAC –15 % / +10 %
-/-
50/60 Hz
50/60 Hz
- 115 VAC –20 % / +10 %
(optional)
-/-
50/60 Hz
50/60 Hz
Power consumption
~400 VA
~1000 VA
~1000 VA
Fuse:
- 90…240 VAC
6,3 A slow
-/-
-/-
- 230 VAC
-/-
6,3 A slow
6,3 A slow
- 115 VAC (optional)
-/-
10,0 A slow
10,0 A slow
Serial interface:
- Serial interface type t
RS 485 optoisolated
- Communication protocol
MODBUS RTU (JBUS)
- Serial transfer speed
9600 Baud
Process variables
Test specimen holder:
- Bore / Depth
60 mm / 163 mm
Housing dimensions:
- Width
- Height
- Depth
~210 mm
~380+50 mm
~300 mm
~147 mm
~330+68 mm
~269 mm
~147 mm
~330+68 mm
~269 mm
Weight
~13,0 kg
~7,5 kg
~7,5 kg
Sensor basket working depth
150 mm