background image

H49.0.02.6C-02

Operating Manual MH 3750

Page 6

6/11

R

pos

T

( )

100 1

3.908310

3

T

+

5.77510

7

T

2

(

)

:=

4 Special Functions

4.1 Display Resolution 

Standard setting: 'Auto', i.e. the device automatically switches over to the optimum resolution between .01° and 0.01°.
If temperatures to be measured are near the switching threshold, a fixed resolution may be better, e.g. for easy 
recording. In such a case please select the optimum resolution manually.

4.2 User Sensor Curve ('Lin USEr')

By means of this function besides the standard conversion of resistance to temperature following EN60751(Lin E.751) 
also other curves can be used. The user sensor curve can be read and edited by the configuration software 
MHKONFIG. The standard setting ex works is also set to the EN60751 data. The curve is defined by a table with two 
columns (input resistance[Ohm]/output temperature [°C]) with 50 rows.

Info: The sensor curve following EN60751 uses the international temperature scale ITS90 and following formulas:

Temperatures <0°C:

Temperatures >=0°C:

4.3 Zero Displacement ('Offset')

A zero displacement can be carried out for the measured temperature:

temperature displayed = temperature measured  -  offset

Standard setting: 'off' = 0.0°, i.e. no zero displacement will be carried out. Together with the scale correction (see 
below) this factor is mainly used to compensate for sensor deviations. Unless the factor is set to 'off', the offset arrow in 
the display shows an active zero displacement.

4.4 Scale Correction ('Scale')

The scale of the measuring can be influenced by this setting (factor is in %):

displayed temperature[°C] = measured temperature[°C] * (1+Scal/100)

respectively displayed temperature[°F] = (measured temperature [°F]-32°F) * (1+Scal/100) + 32°F

Standard setting: 'off' =0.000, i.e. temperature is nor corrected. Together with the zero displacement (see above) this 
factor is mainly used to compensate for sensor deviations.
Unless the factor is set to 'off', the Corr arrow in the display shows an active scale correction.

4.5 Output

The output can be used as serial interface (for GRS3100 or GRS3105 interface adapters) or as analogue output (0-
1V). If none of both is needed, we suggest to switch the output off, because battery life then is extended.

4.5.1 Interface - Base Address ('Adr.')

By using an electrically isolated interface converter GRS3100 or GRS3105 (accessory) the device can be connected to 
a PC. With the GRS3105 it is possible to connect up to 5 instruments to a single interface (please also refer to 
GRS3105-manual). As a precondition the base addresses of all devices must not be identical. In case several devices 
will be connected via one interface make sure to configure the base addresses accordingly. In order to avoid 
transmission errors, there are several security checks implemented (e.g. CRC).

The following standard software packages are available for data transfer:

§

EBS9M

:

9-channel software to record and display the measuring values

§

EASYControl

:

Universal multi-channel software (EASYBUS-, RS485-, and/or MH3000- operation possible) for 

real-time recording and presentation of measuring data in the ACCESS®-data base format.

In case you want to develop your own software we offer a 

MH3000-development package

 including

-

an universally applicable 32bit Windows functions library ('MH3000.DLL') with documentation that can be used by
all 'serious' programming languages.

-

Programming examples for Visual Basic 6.0™, Delphi 1.0™, Testpoint™, Labview™

Note:

The measuring and range values read via interface are always in the selected 
display unit (°C/°F)!

R

neg

T

( )

100 1

3.908310

3

T

+

5.77510

7

T

2

4.18310

12

T

100

(

)

T

3





:=

SIKA, Dr. Siebert & Kühn GmbH & Co. KG, Struthweg 7-9, 34260 Kaufungen, Germany

 

Tel: +49-5605-803-0, Fax: +49-5605-803-54, email: [email protected], web: www.SIKA.net

Summary of Contents for MH 3750

Page 1: ...ert K hn GmbH Co KG Struthweg 7 9 34260 Kaufungen Germany Tel 49 5605 803 0 Fax 49 5605 803 54 email info SIKA net web www SIKA net Operating Manual Precision Thermometer MH 3750 from Version 1 0 For...

Page 2: ...old to a warm environment condensation may result in a failure of the function In such a case make sure the device temperature has adjusted to the ambient temperature before trying a new start up 3 If...

Page 3: ...nfiguration menu 1 3 Connections 1 Output Operation as interface Connect to optically isolated interface adapter accessory GRS 3100 or GRS3105 Operation as analogue output Connection via suitable cabl...

Page 4: ...ed no interface communication takes place oFF Power off function inactive continuous operation e g mains operation oFF Function of the output No output function lowest power consumption SEr Output is...

Page 5: ...heat loss caused by the probe is not considered When measuring fluids therefore the probe should be emerged sufficiently deep and be stirred continuously When measuring gases the probe should also eme...

Page 6: ...ed Together with the zero displacement see above this factor is mainly used to compensate for sensor deviations Unless the factor is set to off the Corr arrow in the display shows an active scale corr...

Page 7: ...ogue Output Scaling with DAC 0 and DAC 1 With the DAC 0 and DAC 1 values the output can be rapidly scaled to Your efforts Keep in mind not to connect low resistive loads to the output otherwise the ou...

Page 8: ...w or by means of the interface and a PC with GSOFT3050 software Max number of measurings 99 A measuring contains current measuring value at the time of recording time and date of the recording After e...

Page 9: ...lly will be stopped Stopping the recording manually By pressing Store key 6 the recording can be stopped manually Then the following choice appears The selection can be made by 5 key 2 and 6 key 5 Qui...

Page 10: ...nnect sensor to socket sensor cable or device defective return to manufacturer for repair Value exceeding measuring range Check Is the value exceeding the measuring range specified temperature too hig...

Page 11: ...is done by integrated horn display and interface Real time clock Integrated clock with date and year Logger 2 Functions individual value logger Func Stor and cyclic logger Func CYCL Memory Stor 99 da...

Reviews: