6 Fault and System Messages
Display
Meaning
Remedy
low battery voltage, device will continue to work for a short
time
replace battery
If mains operation: wrong voltage
replace power supply, if fault continues to
exist: device damaged
low battery voltage
replace battery
If mains operation: wrong voltage
Check/replace power supply, if fault continues
to exist: device damaged
low battery voltage
replace battery
If mains operation: wrong voltage
Check/replace power supply, if fault continues
to exist: device damaged
system error
Disconnect battery or power supply, wait
some time, re-connect
No display
or
weird display
Device does not
react on keypress
device defective
return to manufacturer for repair
Sensor error: no sensor connected
Connect sensor to socket
----
sensor/cable or device defective
return to manufacturer for repair
Value exceeding measuring range
Check: Is the value exceeding the measuring
range specified? ->temperature too high!
Wrong probe connected
Check probe
Err.1
sensor/cable defective
-> replace
Value below display range
Check: Is the
value below the measuring
range specified? -> temperature too low
!
Wrong probe connected
Check probe
Err.2
sensor/cable defective
-> replace
Err.3
Value exceeding display range
-> set resolution to 0.1° or Auto
Err.4
Value below display range
-> set resolution to 0.1° or Auto
Err.7
system error
return to manufacturer for repair
7 Calibration Services
Calibration certificates can be issued by the factory. For this the device has to be sent to the manufacturer. Just the
manufacturer can check the factory settings and correct them if necessary.