EC mAV 2
Maintenance and cleaning
Technical changes reserved
- 43 -
7.4.2
Voltage Measurement Calibration
1. Press the mA % / V button to select the voltage measurement function, and INPUT, CAL
0, V will appear on the display.
2. Connect the EC mAV 2 to the equipment as shown in Figure 7-4.
3. Set the standard source to an appropriate range.
4. Set the output of the standard source to the value given in the table in § 7.4 and when
the output is stabilized, press the 100%/START button, and SAVE will appear on the dis-
play.
The calibrated point has been stored.
5. Press the STEP/AUTO button and CAL FS will appear on the display.
Repeat the operation of step 4.
Figure 7-4